Description Of Controls - Cantel Medical MEDIVATORS renatron II 100 Series Instruction Manual

Dialyzer reprocessing system
Table of Contents

Advertisement

I. Chemicals Required:
The RENATRON
II 100 Series station is designed to function with RENALIN 100 Cold Sterilant. This
®
solution is described in the RENALIN
A. RENATRON II 100 Series Station Controls (Figure 1)
1. "OFF" switch de-energizes electronic circuitry in the RENATRON II 100 Series. POWER WILL STILL
BE TRANSMITTED TO THE MACHINE UNLESS THE MACHINE IS DISCONNECTED FROM THE
ELECTRICAL POWER SUPPLY.
2. "ON" switch energizes electronic circuitry. Sets RENATRON II 100 Series station in standby mode (i.e.,
"volume" display reads ["SEL"] and "PROGRAM" step display indicates [– –]).
3. "START PROCESS" switch initiates dialyzer reprocessing only when a priming volume has been set and
a dialyzer reprocessing program selected i.e., "00," "CH," "HF," "PC". Note: "PC" program can only be
accessed in the automatic mode controlled by the computer. "START PROCESS" also allows user to repeat
most program steps after an alarm situation in a process cycle.
4. "START SANITIZE" switch initiates a machine cleaning cycle and repeats most program steps after an alarm
situation in sanitize cycle.
NOTE – Sanitize can only be initiated if reprocessing program "00," "CH" or "HF" is being displayed in the
"step display," not from the standby [– –] or pre-clean [PC] mode and the volume display reads ["SEL"].
5. "MUTE" switch mutes audio alarm.
6. "RESET" switch returns machine to last program mode used and "VOLUME" display to "SEL". Reset must be
used at the end of a cycle or in an alarm state (locked out or disabled at all other times) to guarantee automatic
transfer of reprocessing data to the computer.
7. "DISPLAY" switch allows operator to blank the display from viewing the actual program steps.
8. "HOLD TO SET" switch when touched, activates "VOLUME" display. Allows operator to view displayed
value of minimum acceptable blood compartment reference volume "SET" into the machine manually or
automatically by the computer.
9. "SET" knob used in conjunction with the "HOLD TO SET" switch allows the operator to set the minimum
acceptable blood compartment reference volume prescribed by the attending physician in the manual mode
of operation.
B. Interface Module Controls:
1. "BAR CODE SCANNER" allows the operator to automatically enter dialyzer and RENATRON II 100 Series
station serial number information into the computer by simply scanning the bar code labels on the dialyzer,
the RENATRON II 100 Series serial number barcode label, and the user authentication label.
2. "Power ON/OFF" switch is located at the back of the Interface Module where the power cord enters the unit.
It allows electrical power to the Interface Module to be switched on and off.
C. RENATRON II 100 Series Station Indicators and Error Messages:
1. "VOLUME DISPLAY" indicates:
a. Blood compartment volume test measurement; or
b. Blood compartment reference volume
2. "PROGRAM STEP" is a display that allows the operator to view the actual program step number and programmed
mode of operation. Display is deactivated/activated by sequentially touching "DISPLAY" switch.
100 Cold Sterilant Instructions for Use (P/N 50090-679).
®
-11-