Cantel Medical MEDIVATORS renatron II 100 Series Instruction Manual page 7

Dialyzer reprocessing system
Table of Contents

Advertisement

CAUTION – IN THE EVENT A DIALYZER HAS BEEN LEFT ON THE MACHINE AND DOUBT EXISTS
WHETHER VENT ACTION HAS OCCURRED, (I.E. POWER FAILURE TO RENATRON II 100 SERIES)
THE DIALYZER MUST BE REMOVED CAREFULLY. REMOVE THE DIALYZER AFTER WRAPPING ALL
CONNECTIONS WITH DISPOSABLE ABSORBENT MATERIAL. PRESSURE BUILD UP MAY CAUSE FLUIDS
TO SPRAY.
The RENATRON
II 100 Series station(s) sit on a single or dual unit drip tray that will catch fluid drippage from
®
the machine(s). A water pressure gauge block, with one or two outlets as appropriate, is attached to the drip tray.
The pressure gauge displays the pressure of the water supplied to the RENATRON II 100 Series station(s). The
steps of the reprocessing process are given below:
1. Performed by Operator
a. Operator connects the dialysate outlet port on the dialyzer to the connector mounted on the front panel
marked "DIALYSATE OUTLET SANITIZE INTERLOCK." This connection acts as the dialyzer holding
fixture.
b. Operator connects the system's "ARTERIAL INLET," "VENOUS OUTLET," and "DIALYSATE INLET"
lines to the dialyzer.
c. In the manual operation mode, the operator sets the blood compartment reference volume by touching
the "HOLD TO SET" switch and turning the "SET" knob until the desired value is displayed in the
"VOLUME" display. The value for the blood compartment reference volume must be determined by the
attending physician for the type and size of the dialyzer to be reprocessed.* The operator also selects the
appropriate reprocessing program mode by pressing the "MUTE" and "RESET" switches.
d. In the automatic operation mode, the operator scans the bar code label on the dialyzer, the serial number
bar code label on the RENATRON II 100 Series on which the dialyzer is to be reprocessed and their
authentication label. The computer will then automatically set the appropriate RENATRON II 100 Series
priming volume and program mode (i.e., "00," "CH," "HF," "PC") for the dialyzer to be reprocessed.
* NOTE – It is recommended that the operator set the reference volume of the dialyzer to be reprocessed to
80% of the new dialyzer priming volume. New dialyzer measured priming volume may differ from advertised
volume by as much as 20%. It is recommended that the operator measure new dialyzer volume with a known
calibrated RENATRON II 100 Series to determine actual new dialyzer priming volume.
Document
Code
20601-183
(00)
20601-184
(CH)
20601-185
(HF)
20601-186
(PC)
2. Automatic Reprocessing Steps
a. After the "START PROCESS" switch is pressed to initiate the reprocessing cycle, the RENATRON
II 100 Series will reprocess the dialyzer according to the steps outlined in the appropriate Operation
Format and Program Outline document.
* NOTE – The RENATRON II 100 Series machine should be sanitized a minimum of ONCE PER DAY.
The machine can automatically sanitize itself when the "Volume" display indicates "SEL" and program mode "00,"
"CH," or "HF" are displayed in the "PROGRAM STEP" display. The operator touches the "START SANITIZE" switch
to start the sanitize cycle. The steps of the sanitize cycle are given below:
b. Operator first visually verifies that the "RENALIN
unexpired RENALIN 100.
NOTE – If an insufficient amount of RENALIN 100 exists to complete one full "SANITIZE" cycle, the machine
will automatically alarm. The message "ADD CHEMICAL" will appear on the front panel and remain visible
until the alarm situation has been corrected and the "START SANITIZE" switch has been pressed.
c. Operator connects the system's "ARTERIAL INLET" and "VENOUS OUTLET" lines to the calibration cell
and connects the "DIALYSATE INLET" line to the "DIALYSATE OUTLET - SANITIZE INTERLOCK"
connector using the male/male fitting attached to the front of the machine.
d. Operator touches the "ON" switch, selects a program by pressing the "MUTE" and "RESET" switches then
touches the "START SANITIZE" switch which activates the sanitize cycle. The machine will automatically
fill and flush all internal blood contact surfaces and the dialysate circuit with Proportioned RENALIN 100
Solution. The machine then enters "PROGRAM STEP" 83, at which time the "SANITIZE COMPLETE"
message will be displayed and a 3 second continuous tone will sound.
Program
Low Flux Dialyzer
Mid-Range/High Efficiency and
Hemophan Type Dialyzers
High-Range/High Flux Dialyzers and
Hemofilters
Pre-clean Program for all Dialyzers
100 CONTAINER" has at least one inch of
®
Kuf
Time
< 8
8 min.
8 - 15
10 min.
15+
10 min.
ALL
4 min.
-7-
Instruction Manual