THE PROPORTIONED RENALIN
REPROCESSED DIALYZER MUST BE ADEQUATELY AND THOROUGHLY RINSED
OUT OF THE DIALYZER PRIOR TO ITS CLINICAL USE. THIS SHOULD BE DONE
IN ACCORDANCE WITH THE INSTRUCTIONS SET FORTH IN THE RENALIN 100
COLD STERILANT INSTRUCTIONS FOR USE (PART #50090-679).
B. Start Up
1. Prior to the first use of day, wipe off the front panel of the RENATRON
disposable wipe saturated with fresh 1% RENALIN 100 Solution or full strength ACTRIL
Discard the used wipe.
2. Verify that the Calibration Cell is attached to the RENATRON II 100 Series Station.
3. Turn on the water supply and check static water pressure reading (should be between 20-55 psi).
4. Verify that there is sufficient RENALIN 100 in the container to reprocess the dialyzer that will be
installed on the RENATRON II 100 Series station in step D, below and that the RENALIN 100 Cold
Sterilant is not expired. Approximately 1" of RENALIN 100 Cold Sterilant is required. Verify that the
uptake hose is connected to the container.
5. Touch the "ON" switch and select program mode "00".
a. To change the program mode, simultaneously press "MUTE/ALARM" and "RESET".
b. "START PROCESS" will not initiate if "SEL" is displayed in the upper window. To clear, press
"HOLD TO SET" and turn "SET" knob to 67.
6. Check dynamic water pressure. In step 4, the dynamic pressure should optimally be between 30-35 psi.
7. Verify that the Calibration Cell volume displayed is within ± 3 ml of the recorded reference volume.
8. Remove the Calibration Cell.
9. Press "RESET."
Note – When calibrating the RENATRON II 100 using the RENALOG RM Dialyzer Reprocessing
Management System, follow the calibration procedure outlined in the RENALOG RM User Guide (PN
50090-859).
C. Remove the Dialyzer from the Extracorporeal Blood Circuit using standard facility protocols.
Note – Prior to removing the dialyzer from the extracorporeal circuit, the operator must ensure that all
identification and information on the dialyzer is legible and current before proceeding.
THE OPERATOR SHOULD WEAR PROTECTIVE EYEWEAR, GLOVES, AND OTHER
PROTECTIVE CLOTHING. FOLLOW UNIVERSAL SAFETY PRECAUTIONS FOR
DEALING WITH BLOOD CONTAMINATED DEVICES.
CAUTION – THE TIME INTERVAL BETWEEN REMOVING THE DIALYZER FROM THE EXTRACORPOREAL
CIRCUIT AND REPROCESSING IT SHOULD BE KEPT WITHIN THE LIMITS PRESCRIBED BY THE PHYSICIAN.
THE SUGGESTED MAXIMUM TIME INTERVAL IS TWO (2) HOURS.
D. Installation of the Dialyzer on the RENATRON II 100 Series station.
1. Remove the blood port cap from the venous blood port of the dialyzer.
2. Wipe the venous blood port with a wipe saturated with 1% RENALIN 100 Cold Sterilant Solution or full
strength Actril Cold Sterilant and attach a disinfected reprocessing connector.
3. Attach the RENATRON II 100 venous outlet line to the reprocessing connector attached to the venous port of
the dialyzer.
4. Connect the dialyzer dialysate inlet to the RENATRON II 100 dialysate inlet.
5. Connect the dialyzer dialysate outlet to the RENATRON II 100 dialysate outlet/sanitize interlock.
6. Remove the blood port cap from the arterial blood port.
7. Wipe the arterial blood port with a wipe saturated in 1% RENALIN 100 Cold Sterilant Solution or full strength
Actril Cold Sterilant and attach a disinfected reprocessing connector.
8. Attach the RENATRON II 100 Series station arterial inlet line to the reprocessing connector attached to the
arterial blood port of the dialyzer. Dialyzer placement should be vertical to the RENATRON II 100 Series
station.
WARNING
100 SOLUTION CONTAINED IN THE
®
II 100 Series station with a
®
WARNING
-19-
Cold Sterilant.
®
Instruction Manual

Advertisement

Table of Contents
loading

Related Products for Cantel Medical MEDIVATORS renatron II 100 Series

This manual is also suitable for:

Rs 8330Rs 8335

Table of Contents