Troubleshooting

message
caused by
There is a problem with
the meter.
There is an abnormal
Check Strip signal.
Error message could be
caused by a used or wet
test strip.
Poor amount of
blood sample.
There is a problem
with the test strip
Test strip is damaged.
Error message that indicates
that the blood or control
solution sample was applied
before the
appeared on the display.
46
what to do?
Repeat the Check Strip
test 2 or 3 times. Do not
use the meter. Contact
B. Braun representative.
Repeat the test with a
new test strip.
Retest with new test strip
and sufficient blood sam-
ple.
Repeat the test with a
new test strip.
Repeat the test with new
test strip. Apply blood or
control solution only after
sym bol
the
symbol appears
on the display.
message
caused by
No results obtained on the
meter.
The test result is
higher than 600 mg/dL
(33.3 mmol/L).
The test result is
lower than 10 mg/dL
(0.6 mmol/L).
what to do?
The meter memory is
empty until you perform
a first blood glucose mea-
surement.
In case of doubt of
blood test result, check
the meter with control
solution. With normal
result, re-test with blood
sample two or three times.
When "HI" message
persists, please consult
a doctor for assistance.
In case of doubt of
blood test result, check
the meter with control
solution. With normal
result, re-test with blood
sample two or three times.
When "Lo" message
persists, please consult
a doctor for assistance.
47