Summary of Contents for B. Braun Omnitest plus

  • Page 1 Omnitest plus ® Blood Glucose Monitoring System www.omnitest.bbraun.com...
  • Page 2: Table Of Contents

    Important information Content The Omnitest plus system provides a quick and easy way to ® Omnitest plus Blood Glucose Monitoring System ® measure the blood glucose level. Omnitest plus could be used for ® The Meter, Display Segments Quick Reference self-monitoring of blood glucose level by diabetes patients.
  • Page 3: Omnitest ® Plus Blood Glucose Monitoring System

    ® Omnitest plus Control ® Check strip shown at the outer packaging. If anything is missing, control solution please return your system to the place of purchase or (sold separately) contact your nearest B. Braun representative.
  • Page 4 The Meter Display Segments Test strip port Segment Meaning Segment Meaning Data transfer port The battery symbol means “AVG” is displayed when average results are shown. the battery is low and LCD display should be changed soon. Control solution symbol The “DEL”...
  • Page 5: Quick Reference

    Quick Reference CODE? ✔ sec .
  • Page 6 Quick Reference...
  • Page 7: Omnitest ® Plus Blood Glucose Test Strip

    Omnitest plus Blood Glucose Test Strip ® Omnitest plus test strips are only to be used with Omnitest plus Use test strip immediately after removing it from the vial. ® ® ■ meter for accurate analysis of your blood glucose level. Do not use test strips after the expiration date printed on the ■...
  • Page 8: Code The Meter

    Code the Meter Code numbers are used to calibrate the Omnitest plus meter for ® Press button to increase the code number or button to accurate results. You must confirm the code number of the test strip decrease the code number. The number increases or decreases by vial with the number shown on the meter, before using Omnitest ®...
  • Page 9: Battery

    Battery Checking the Meter Display Change the battery when the battery Each time you enter a new battery, all segments of the display will icon appears on the display or when appear for 3 seconds. The numbers displayed later are only meant the meter does not power on by any for internal tests of the electronics.
  • Page 10: Check Strip

    Check Strip Using Control Solution The check strip is used to prove that the Omnitest plus meter is The Omnitest plus Control solution is used to check ® ® <<< working properly. Insert check strip into meter, facing “ ” up. that the blood glucose monitoring system consisting Meter will be automatically turned on.
  • Page 11 Using Control Solution Insert Test Strip Control Solution Test Mode If the control solution is cold, do not use When blood and test strip symbol until the solution has warmed up to appears, press button to mark room temperature. Control solution tests the test as a control solution test in the have to be done in a temperature range meter memory.
  • Page 12 Using Control Solution Apply Control Solution Comparing Control Solution Results Shake the control solution vial gently. If test results fall outside the range, repeat the test. Out-of-range Remove the cap. Squeeze the vial, discard results may be caused by one or more of the following causes: the first drop, and wipe off the dispenser Error in performing the test.
  • Page 13: Set The Meter

    Set the Meter The Omnitest plus meter has 7 functional setting modes Set the Buzzer ® Turn on the meter by pressing button and hold for 3 seconds. buzzer Buzzer icon will appear on the display. Press to select 12 hour/24 hour ON/OFF the buzzer sound.
  • Page 14: Set The Date And Time

    Set the Meter Set the Date and Time Set the Test Result Unit Press button to select the year. With the correct year on Press button to set the test result unit. You should now be at the display, press the button and the month setting will start fla- unit of measurement selection.
  • Page 15: Set The Average Day

    Set the Meter Set the Average Day Set the Alarm button to proceed to set the Average Day. Omnitest plus Omnitest plus has 5 different alarm settings. The bell icon Press ® ® will show 3 different average results and the number of days can will be displayed during the alarm setting.
  • Page 16: Reviewing Your Results

    Reviewing Your Results Deleting Test Results Omnitest plus saves up to 250 test results in built-in memory and ® 3 different average glucose levels are calculated and saved. These values will appear on the display by recalling anytime you like to see To delete any individual test result in them.
  • Page 17: Collecting A Drop Of Blood Sample

    Collecting a Drop of Blood Sample Unscrew the Omnilance tip and insert Next hold the tip in one hand and pull a new Omnican lancet firmly into the on the sliding barrel with the other ® carrier. hand. As the ends are pulled apart, the spring tension increases.
  • Page 18: Testing Your Blood Glucose

    Collecting a Drop of Blood Sample Testing Your Blood Glucose Set aside Omnilance and wait a few seconds for a blood drop Cleanliness: to form. The flow of blood will be helped keeping the hand warm, Wash your hands with warm soapy water. Rinse and dry hands tho- by lowering the hand to waist level, and by gentle massage of the roughly.
  • Page 19 Testing Your Blood Glucose Insert the Test Strip Compare the Code Number Remove test strip from the vial. Recap The meter will automatically turn on and the vial immediately to prevent moisture display the code number. Make sure the from affecting the other strips. Insert a code number on the display matches the test strip.
  • Page 20 Testing Your Blood Glucose Applying Blood Sample After 3 seconds of code confirmation Touch your finger to the tip of Omnitest plus test strip. ® blood icon will be displayed on the screen. Date and time information are shown at the bottom. The blinking reminds you to close the test strip vial cap immediately.
  • Page 21 Testing Your Blood Glucose Accurate Results in Seconds After “beep” sounds, the test will begin The measurement chamber of the test strip auto matically, counting down the numbers will draw automatically the blood of your from 5 to 1 on the display. finger.
  • Page 22 Testing Your Blood Glucose Blood glucose test results are displayed on the Omnitest plus ® Remove Tested Strip by Ejector meter as either milligrams of glucose per deciliter of blood (mg/dL) or millimoles of glucose per liter of blood (mmol/L), depending Slide the ejector button forward to remove the test strip from the upon the test result unit selected (see page 27).
  • Page 23: Maintain Your Blood Glucose Monitor System

    When you use up the lancets or lancing device is damaged, you can purchase them at your nearest pharmacy or B. Braun Do not soak the meter and test strip into water or liquid. Use representative.
  • Page 24: Troubleshooting

    There is an abnormal use the meter. Contact a first blood glucose mea- Check Strip signal. B. Braun representative. surement. Error message could be Repeat the test with a The test result is In case of doubt of caused by a used or wet new test strip.
  • Page 25 10 ~ 40 °C contact your distributor (50 ~ 104 °F) for more than or nearest B. Braun repre- 10 minutes and re-test. sentative. Test does not Poor amount of sample...
  • Page 26: Specifications

    Specifications Test Range 10 ~ 600 mg/dL 0.6 ~ 33.3 mmol/L Symbol Description Reading Time 5 sec. Consult operating instructions Memory Capacity 250 test results Used By ° ° ° ° Operating Temperature C ~ 40 F ~ 104 This product fulfills the requirements of Directive 98/79/EC on in vitro diagnostic medical devices Operating Humidity...
  • Page 27: Transfer Test Results To A Computer

    Specifications Transfer Test Results to a Computer Test Principle: You can transfer test results from the Omnitest plus meter to a com- The enzyme glucose oxidase on the test strip reacts specifically with ® the blood glucose. The current generated is converted and displayed puter, where the data can be further as blood glucose value.
  • Page 28: Blood Glucose Conversion Table

    Blood Glucose Conversion Table Blood Glucose Conversion Table (mmol/L to mg/dL) mmol/L 0.55 mg/dL mmol/L 10.0 10.5 11.0 11.1 12.0 12.5 13.9 mg/dL mmol/L 14.4 15.0 16.0 16.6 17.0 18.0 19.0 20.0 20.8 22.2 23.0 24.0 25.0 26.4 27.7 30.0 33.3 mg/dL...
  • Page 29 If you have any questions about the use of the Omnitest plus ® system, please contact your near- est B. Braun representative or go www.omnitest.bbraun.com B. Braun Melsungen AG Carl-Braun-Straße 1 34212 Melsungen Germany Tel ++49 (0) 56 61 71-33 99 Fax ++49 (0) 56 61 71-35 50 www.bbraun.com...

Table of Contents