ImageXpress Pico Automated Cell Imaging System Calibration Kit Guide
4. Note the status of the filter cube, which is indicated in the tile.
5. In the tile for the filter cube you want to calibrate, click
6. Follow the on-screen instructions to complete the calibration.
30
If the tile indicates Not Calibrated, continue to the next step to perform the
calibration procedure.
If the tile indicates Calibrated, you typically do not need to perform the calibration
procedure and you can skip this procedure.
Note: An exception to this is when you replace a filter cube with the same type of
filter cube. In this case, you should continue to the next step and perform the
calibration procedure regardless of the label in the tile.
Tip:
In the Choose Slide Holder drop-down list box, select 4 Slide Holder.
Insert the slides face down in the slide holder (that is, printed side down).
When each step completes, click Next to continue to the next step.
Filter Cube Calibration.
5063669 E