Fluke Calibration 6050H Final Assembly Manual page 2

Setting the hard cut-out

Advertisement

Checking the controller CTO
Before we set the Internal
1
Hard Cut-Out, we must first
see what the Hard Cut-Out
(CTO) is set to.
6050HCTO
Press the "SET" and "EXIT"
2
keys together to enter into the
Secondary Menu. The heater
duty cycle will be displayed.
From Here, press the "SET"
key until "PrObE" is displayed.
Rev 001
Press "UP" key until "CAL" is
3
displayed.