Der Inhalt dieser Produktbeilage wird regelmäßig aktualisiert und ist beim Einsatz des Produktes unbedingt zu beachten.
Sie fi nden diese Produktbeilage auch im Download-Bereich unserer Website unter www.fi or-gentz.de/downloads.
Date printed: 2018-11
GB
This manual is regularly updated and should be strictly followed.
You can also fi nd this manual in the download section on our website at www.fi or-gentz.com/downloads.
Nota:
IT
Il presente manuale è disponibile anche in lingua italiana nella sezione download del nostro sito www.fi or-gentz.com/downloads.
Systemknöchelgelenk
System Ankle Joint
Seite 2
Page 18

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Neuro Swing and is the answer not in the manual?

Questions and answers

Summary of Contents for FIOR & GENTZ Neuro Swing

  • Page 1 Systemknöchelgelenk System Ankle Joint Druckdatum: 2018-11 Der Inhalt dieser Produktbeilage wird regelmäßig aktualisiert und ist beim Einsatz des Produktes unbedingt zu beachten. Seite 2 Sie fi nden diese Produktbeilage auch im Download-Bereich unserer Website unter www.fi or-gentz.de/downloads. Date printed: 2018-11 This manual is regularly updated and should be strictly followed.
  • Page 2: Table Of Contents

    2. Warranty 3. Safety Information 4. Safety Instructions 4.1 Classification of the Safety Instructions 4.2 All Instructions for a Safe Handling of the NEURO SWING System Ankle Joint 5. Application 6. Joint Function 7. Scope of Delivery 8. Load Capacity 9.
  • Page 3: Declaration Of Conformity

    1. Declaration of Conformity We declare that our medical devices as well as our accessories for medical devices are in conformity with the requirements of the Medical Devices Directive 93/42/EEC. Therefore, the FIOR & GENTZ products bear the CE marking. 2.
  • Page 4: All Instructions For A Safe Handling Of The Neuro Swing System Ankle Joint

    4.2 All Instructions for a Safe Handling of the NEURO SWING System Ankle Joint DANGER Potential Traffic Accident Due to Limited Driving Ability Advise the patient to gather information about all safety and security issues before driving a motor vehicle with orthosis. The patient should be able to drive a motor vehicle safely.
  • Page 5 WARNING Jeopardising the Therapy Goal Due to Lack of Free Movement Check if the system joint moves freely in order to avoid restrictions of the joint function. Use suitable sliding washers according to the information in this manual. WARNING Jeopardising the Therapy Goal Due to Incorrectly Filed System Stirrup Do not file the system stirrup too far.
  • Page 6: Application

    6. Joint Function The basic function of the NEURO SWING system ankle joint is to dynamically bring the patient from a bent into an upright position as well as to stabilise the patient when walking and standing by balancing the body. Spring units of different strengths, that are mounted to the system ankle joint, offer the maximum range of motion in plantar flexion and dorsiflexion.
  • Page 7: Load Capacity

    8. Load Capacity You can find all relevant patient data for your selected system joint in the configuration you performed with the Orthosis Configurator and/or on your completed orthotic treatment sheet. 9. Tools for Assembling the System Joint System Width 10mm 12mm 14mm...
  • Page 8: Assembly/Lamination Dummies

    “5 – Laminating the AFO” of the online tutorial “AFO in Joint Lamination Technique” on our website www.fior-gentz.com. 11.1 Mounting the Cover Plate Start the assembly of the NEURO SWING system ankle joint by mounting the cover plate. Proceed as follows: fi g. 8 1.
  • Page 9: Mounting The System Stirrup

    Mount the system stirrup corresponding to the system width and production technique to the NEURO SWING system ankle joint. Proceed as follows: 1. Grease the axle bore of the system stirrup as well as the friction surfaces of the bearing nut and the second sliding washer with the delivered orthosis joint grease.
  • Page 10: Checking The System Joint's Free Movement

    11.3 Checking the System Joint's Free Movement If the system joint does not move freely, some of the spring force will be wasted by overcoming the friction forces. As a consequence, the joint function is restricted and the treatment goal might not be achieved. Check if the system joint moves freely.
  • Page 11: Adjustment Options On The Orthosis

    12. Adjustment Options on the Orthosis The NEURO SWING system ankle joint can be individually adjusted to the patient's pathological gait with the help of three adjustment options (fig. 22): - adjustable alignment using the alignment screw (1); - adjustable range of motion using the motion limiting screw (2);...
  • Page 12: Variable Spring Force

    12.3 Variable Spring Force The spring force in plantar flexion and dorsiflexion can be changed with- out great effort by using spring units of different strengths. There are five spring units with spring forces ranging from normal to extra strong (fig. 27).
  • Page 13: Maintenance

    14. Maintenance When mounting and demounting the system joint too often, the bars of the cover plate can deform due to the elasticity of the material and thus cause lateral play in the system joint. In this case, the acting force of the spring units can increase the play.
  • Page 14: Replacing The Sliding Washers

    Boring and Reaming Measurements [mm] Repair Repair Ø Measure- Ø Measure- Art. No. System Bearing Nut Bushing Bushing ment for ment for Repair Width Outer Ø Inner Ø Outer Ø Boring Reaming Bushing 10mm 7.0 H7 BP0706-L020 12mm 8.0 H7 BR0807-L020 14mm 9.6 H7...
  • Page 15: Spare Parts

    16. Spare Parts fi g. 40 Article Number for System Width Item 10mm 12mm 14mm 16mm 20mm Description bearing nut SB6039-L0530 SB7049-L0540 SB8559-L0580 SB9669-L0710 SB1069-L0810 upper part, SF0310-ST SF0311-ST SF0312-ST SF0313-ST SF0315-ST straight, steel upper part, SF0311-TI SF0312-TI SF0313-TI SF0315-TI straight, titanium upper part, bent SF0330-ST...
  • Page 16 Article Number for System Width Item 10mm 12mm 14mm 16mm 20mm Description countersunk SC1403-L08 SC1404-L10 SC1405-L11 SC1405-L12 SC1405-L12 flat head screw, hexalobular socket countersunk flat head screw, SC1403-L08 SC1404-L10 SC1405-L11 SC1405-L12 SC1406-L12 hexalobular socket (axle screw) O-ring for securing VE3771-045/10 VE3771-06/10 VE3771-07/10 VE3771-08/10 VE3771-11/10...
  • Page 17: Information For The Treatment Documentation

    17. Informationen für die Versorgungsdokumentation/Information for the Treatment Documentation Bitte heften Sie diese Produktbeilage zu Ihrer Versorgungsdokumentation! Add this manual to your treatment documentation! Patientendaten/Patient´s Data Name Name Straße Address PLZ, Wohnort Postcode, City Telefon privat Home Telephone Telefon geschäftlich Telephone at Work Kostenträger Insurance...
  • Page 18 Beinseite Leg Side links/left rechts/right Montierte Gleitscheiben-Version Mounted Sliding Washer Version 1. GS ____________ - _________ 2. GS ____________ - _________ Der Qualitätsstandard der Firma FIOR & GENTZ ist durch eine unabhängige Zertifi zierungsgesellschaft nach den internationalen Normen ISO 9001 und ISO 13485 geprüft und bescheinigt worden. The quality standard of the FIOR &...