Protection; Custom Linearization; User Calibration - gefran 40B 48 User Manual

Position, force and pressure indicator-interceptor with input for strain-gauge and potentiometer
Table of Contents

Advertisement

if U.C. = 1
Note: between the calibration phases some seconds could be requested to elaborate data.
81651C_MHW_40B48-96_0308_ENG
P r
Protection code
+4 to disable In and Ou pages
+8 to disable Cf page
+16 to enable maintenance of reset memory at shutdown
+32 base configuration (the following parameters will not be displayed:
In: Ft, Fd, Of, L_L, H_L
Ou: On [forced to no. outputs present], rE)

• Custom Linearization

Custom Linearization of
L n
main input
enabled only if t.P = Linear Custom (only for version SW.2.0x)
Step 0
(beginning of scale value)
.......
Step 32
(full scale value)

• User Calibration

U. C.
if U.C. = 2
Minimum
calibration (*)
Maximum
calibration (*)
∆ ∇
(*) Press keys
to
calibrate the analog
output
P.V.
.......

• Protection

Value
Displayed
Modifiable
parameters
parameters
0
o.1, o.2, o.3
o.1, o.2, o.3
1
o.1, o.2
o.1, o.2
2
o.1
o.1
3
o.1
none
Display limits
(-1999 to 9999 for 4 digit display)
the n step value corresponds to input:
mV beginning scale + n *
mV
mV = (mV full scale - mV beginning scale) / 32
Display limits
(-1999 to 9999 for 4 digit display)
U.C.
CPU B function
1
analog retransmission
output
2
potentiometer
3
strain-gauge
positive polarization
4
strain-gauge
symmetrical polarization
if U.C. = 3, 4
Zero acquisition phase,
potentiometer with cursor
in
minimum
voltage
position
Maximum acquisition
phase, potentiometer
with cursor in maximum
voltage position
Process variable
(LEVEL 1)
Tare zero acquisition phase with strain-
gauge discharged (no weight or
pressure)
Acquisition phase for automatic definition
of sensitivity and full scale, load strain-
gauge with sample reference or
automatic activation of configured output
to check 6-lead sensor (only for version
SW 2.0x). 80% f.s. is set as default. The
value can be changed to set the value in
engineering units corresponding to the
sample used.
Zero recalculation phase, unload strain-
gauge by removing sample reference.
5