Specifications - Spectrum Technologies Field Scout 2266FS Product Manual

Direct soil ec meter
Table of Contents

Advertisement

Conductivity
Range
Resolution
Relative accuracy
Normalization temp.
Temperature co-efficient
Calibration points
TDS
Range
Resolution
Relative accuracy
Calibration points
TDS factor
Salinity
Range
Resolution
Relative accuracy
Calibration points
Temperature
Range
Temperature resolution
Temperature accuracy
General
Display Graphics
Auto off
Power requirement
Battery life
Water proofing
Environmental operating conditions
Ambient temperature
Relative humidity
Storage temperature
Storage humidity
Probe Dimensions

Specifications

0.0 to 200.0 μS, 200 to 2000 μS,
2.00 to 20.00 mS
0.1 μS, 1 μS, 0.01 mS
±1% full scale
25.0ºC (77ºF)
0.0% to 10.0%
Up to 3 points
0.0 to 100.0 ppm, 100 to 1000 ppm,
0.10 to 10.00 ppt (TDS Factor 0.5)
0.1 ppm, 1 ppm, 0.01 ppt
±1% full scale
Up to 3 points
0.40 to 1.00 (selectable)
0.00 to 10.00 ppt
0.10 ppt
±1% full scale
1
0 to 60°C (32.0 to 140.0°F)
0.1°C / 0.1°F
From 0 to 50ºC (±0.5ºC / ±0.9ºF + 1 LSD);
from 50 to 60ºC (±1.0ºC / ±1.8ºF + 1 LSD)
Dot matrix 80 x 100 pixel
8.5 minutes (from last key press)
Four AAA 1.5 V batteries
>150 hours
IP67
5 to 45°C / 41 to 113°F
5% to 85% noncondensing
–20 to 60°C / –4 to 140°F
5% to 85% noncondensing
2266FS
Length 7.7 in (19.5 cm)
Diameter 0.30 in (0.8 cm)
2266FSTP
Length 24 in (61 cm)
Diameter 0.35 in (0.9 cm)
17

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Field Scout 2266FS and is the answer not in the manual?

This manual is also suitable for:

Field scout 2266fstp

Table of Contents