Power ..................................................................... Lithium-Ion rechargeable, 7.2 V, 31 Wh battery, 4300 mAh
Battery Charger ..................................................... 100 V to 240 V, 50/60 Hz input, 15 V/2.0 A output. For best performance, the battery
Battery Life ............................................................ 9 hours (minimum), 100 NIBP cycles typical
Weight .................................................................... 1.81 kg (4 lb)
Wireless Radio
Frequency Range ............................................... 2412 MHz to 2462 MHz
Output Power ...................................................... <1 mW
Safety ..................................................................... IEC 61010-1: Overvoltage Category II, Pollution Degree 2
Electromagnetic Compatibility (EMC)
International ........................................................ IEC 61326-1: Portable Electromagnetic Environment CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the
internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage power
supply network that supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic
compatibility in other environments due to conducted and radiated disturbances.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object. The equipment
may not meet the immunity requirements of this standard when test leads and/or test probes are connected.
Korea (KCC) ....................................................... Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it.
This equipment is intended for use in business environments and not to be used in homes.