LS-1310/1550
Laser Source

Specifications

Accuracy is specified for a period of one year after
calibration, at 18 °C to 28 °C (64 °F to 82 °F) with
relative humidity to 90%. Specifications are shown
in Table 1.
Table 1. Specifications
Emitter Type
Laser
Classification
Safety
Output
Wavelengths
Output Connector
Output Power
(CW)
Output Level
Stability
Temperature
Range
Power Source
Size
Weight
Battery Life
Indicators

Cleaning the Optical Connector

Most problems with optical sources result from
contaminated connectors. Therefore, always clean
the connector before troubleshooting or calibration.
To clean the source connector, wipe it gently with
an optical-grade tissue or swab dampened with
optical-grade alcohol. To remove loose dirt and
dust from the connector, use filtered, compressed
air.
Always cover the connector with a dust cap when
the unit is not in use.
2
Laser
Class 1
Complies with ANSI/ISA S82.01-
1994, CSA C22.2 No. 1010.1-92,
EN61010.1: 1993
1310 nm ±20 nm
1550 nm ±30 nm
Singlemode ST
-10 dBm (100 µW) adjustable
At 25 °C after a 15-minute
warm-up:
±0.2 dB over 1 hour
±0.4 dB over 8 hours
Operating: 0 °C to 40 °C
Storage: -10 °C to +60 °C
9 V battery
17.4 cm x 7.6 cm x 3.8 cm
(6.9 in x 3.0 in x 1.5 in)
266 g (9.38 oz)
16 hours typical (alkaline)
SOURCE ACTIVE (power on)
LED
LOW BATTERY LED

Required Equipment

The following equipment is required for
performance tests and calibration adjustments:
Fiber optic power meter calibrated at 1310 nm
and 1550 nm
0-10 V dc adjustable power supply with a
minimum output of 500 mA and a current-limit
adjustment
Fluke 187 DMM (digital multimeter),
equivalent, or better
20 MHz digital or analog oscilloscope,
equivalent, or better
Optical power meter (OPM) traceable to local
national standards
ST/ST singlemode fiber patch cable
Using a multimode fiber patch cable will
cause calibration errors.

Performance Tests

Use the performance tests to confirm that the
source is working properly. If the source fails any
of these tests, clean the fiber connectors, verify that
the connections are solid, then retest the laser
source. If the source still fails a test, it needs
calibration adjustments or repair.
1. Turn on the optical power meter (OPM).
2. Set the laser source's power switch to 1310 nm.
Verify that the
LOW BATTERY
briefly before the
SOURCE ACTIVE
on. Let the laser source stabilize for 5 minutes.
3. Using a singlemode fiber patch cable, connect
the laser source to the OPM.
4. Set the OPM to measure 1310 nm. Set the laser
source's
switch to
CW/MOD
5. Verify that the power reading is -10.00 dBm
±0.5 dB.
6. Set the laser source's CW/MOD switch to
MOD. Verify that the power reading increases
by 0.6 dBm +0.3 dB.
7. Set the laser source's
set the power switch to 1550 nm. Set the OPM
to measure 1550 nm.
8. Repeat steps 5, 6, and 7.
9. Turn off the laser source. Disconnect the fiber
patch cable; then install the dust caps on the
source.
Note
LED flashes
LED turns
CW.
CW/MOD
switch to
CW
;