Signal Connections
To null the static voltage offset of the system, including the bridge, use the following procedure:
1. Configure and connect your bridge to the selected channel.
2. Read the channel output.
3. While monitoring the output, rotate the trimming potentiometer wiper with a flathead
screwdriver until you reach 0 V.
You have nulled your system offset and are ready for a measurement.

Nulling Range Adjustment

The nulling range of the offset nulling circuitry is approximately ±5 mV referred to input (RTI),
assuming an excitation voltage of 2.5 V. The nulling circuit of each channel has a resistor which
sets this nulling range. You can change the nulling range of the offset nulling circuitry for each
channel by replacing its nulling resistor with a resistor of another value. Therefore, you can mix
your ranges to accommodate each channel requirement. Table 3-7 lists the nulling resistors and
their corresponding channels.

Table 3-7. Nulling Resistor and Corresponding Channel

The value of all the nulling resistors on the SC-2043-SG is 47 kΩ. Notice that these resistors are
socketed for easy replacement. These sockets best fit a 1/4 W resistor lead size.
If you want to change the nulling range of any channel, use the following formula to determine
the nulling resistor value you need to achieve your desired nulling range:
R
EX
null
V
desired nulling range
where:
R
is the nulling resistor value,
null
EX is the excitation voltage (factory-adjusted to 2.5 V),
V
desired nulling range
SC-2043-SG User Manual
Channel
0
1
2
3
4
5
6
7
100
RTI
RTI is the desired nulling range, referred to input.
Nulling Resistor
R68
R53
R34
R18
R69
R55
R35
R19