14. Technical specifications

Model No.
Measurement method
Measurement range (blood
pressure measurement)
ECG bandwidth/sample rate
ECG Heart Rate range
Display accuracy
Measurement inaccuracy
Memory
Dimensions
Weight
Cuff size
BM 93
Oscillometric, non-invasive blood pressure measurement on the
upper arm.
Single-channel ECG in freely selectable front positions / ECG sig-
nal ground (earth) referenced.
Cuff pressure 0 – 300 mmHg,
systolic 50 – 280 mmHg,
diastolic 40 – 150 mmHg,
Pulse 40 –199 beats/minute
0.05 to 40 Hz/125 Hz
5 –199 beats/minute
Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5% of the value shown
Max. permissible standard deviation according to clinical testing:
Systolic 8 mmHg/diastolic 8 mmHg
2x 60 memory spaces
Main unit: L 128 mm x W 128 mm x H 40 mm
Main unit: Approximately 300 g (without batteries, with cuff)
23 to 43 cm
55