Storage; Installation And Initial Start-Up; Safety Notes For Installation And Initial Start-Up; Preparation - GCE druva MPLH0XS0 Instructions For Use Manual

Table of Contents

Advertisement

Available languages

Available languages

5.4. STORAGE

Store the packages in the following conditions:
Do not store outdoors
Store in a dry and dust-free location
Do not expose to aggressive media
Protect from sunlight radiation
Avoid mechanical jolts
EN
Storage temperature: 15 to 35 °C
Relative humidity: max. 60 %
If storing for longer than 3 months, regularly inspect the general condition of all parts and the packaging.
If necessary re-apply or renew the rust-proofi ng
TIPS AND RECOMMENDATIONS!
Some packages may bear labels with storage information that extends beyond these requirements. These
notes should be observed accordingly.
6. INSTALLATION AND INITIAL START UP
6.1. SAFETY NOTES FOR INSTALLATION AND INITIAL START UP
STAFF
Installation and initial start-up of the system may only be performed by qualifi ed sta .

6.2. PREPARATION

UNPACKING
The system components should be removed from their packaging carefully and prudently.
Additional protective packaging should also be removed.
Check all components of damages from transport
DEPRESSURIZE
Depressurize components and purge with inert gas if necessary
Cut tubing with special tool; avoid contaminations (dirt, cuttings, etc.)
Check perfect condition of components and purity of connections

6.3. INSTALLATION

The back plate of the manifold is mounted at the height of 164 cm. The front plate is suspend in the back plate
and secured with the delivered screw below in the middle. Connect grounding on ground plate.
The manifold is mounted with compression fi ttings at process outlet, relief valve, purge inlets and purge
outlets, if existent connect safety and/or check valve.
First remove the plastic caps from all connections.
The piping has to be inserting completely into the compression fi tting.
Than screw the nut hand tight. After that screw with a jaw spanner 1 ¼ turns tight.
Connect the relief tubing the same way. It is not allowed to connect relief and purge tubing. They have to go
separately and safe to the outside, also the safety valve tubing.
To connect the gas cylinder to the manifold coils and fl exible hoses are suitable (available accessories)
The stainless steel coils or fl exible hose is always delivered separately. The correct allocation has to be
proved. At the connection nut you can see the type of cylinder connection. There is only one design of coil/
hoses to connect it at the right and left side of manifold.
To mount the coil/ hose please remove plastic caps from the thread. Make sure that the gasket, which is
scope of delivery, is inserted. Connect the nut with the inlet of the manifold hand tight and afterwards screw
with a jaw spanner.
To connect the coil/hose with the gas cylinder, the thread of the cylinder valve and of the nut need to be in
perfect condition. Any time you connect new gas cylinder, use new cylinder connection.
Only coils and hoses from manufacturer suitable for used gases have to be used. Check gasket or correct
position in connection thread of coil/hose. Using a spanner extension is not allowed, it can cause damage of
thread and gasket and leads to leakage.
26/68