2.5 Battery Replacement

The "LO" annunciator on the LCD alerts when battery power is running
low.
Caution: Power off the meter before changing battery.

2.6 Electrode Information

The meter includes a probe with a BNC connector
having a nominal cell constant of k = 1.0, and a built-in
temperature sensor. The Ultem-body housing has good
chemical resistant properties. The probe design off ers fast
temperature response and reduces air entrapment, ensuring
accurate, repeatable, and stable readings.
The wettable materials of the probe include:
Polyetherimide (Ultem) – protective probe guard
1.
2.
Polybutylterphalate (Valox) – sensor housing
Stainless Steel (SS 304) – 2 steel bands
3.
The protective probe guard can be removed temporarily
for maintenance but must be reattached before
measurement and calibration. Erroneous results will occur
during calibration and measurement if the probe guard is
not used.
Always immerse the probe beyond upper steel band for
best results. Use the fi ll line on the outside of the probe
guard for reference.
LO
"LO" Battery Condition
S
μ
9