Taking A Measurement - Schaller Messtechnik humimeter FS1 Operating Manual

Grain moisture meter
Table of Contents

Advertisement

5.2

Taking a measurement

To do so:
The device has to have nearly the same temperature
than the product being measured.
1.
Fill the measuring cup with the filling quantity displayed
in the product type name (+/- 1.0 g) (figure 12)
(see
"6. Product
2.
Now slowly and evenly fill the measuring chamber of
the device with the material being measured (figure 13).
»
For the filling, no funnel or similar device may be used.
3.
The device will now display the moisture content
(figure 14).
»
The displayed value flashes when the moisture
content exceeds the measuring range of the se-
lected product type (figure 15). A flashing value
signals a decreasing accuracy of the measure-
ment. The measuring range is dependent on the
product type (see
»
Once the reading has been taken, the measuring
values can be frozen on the display (see
function - Freezing the displayed
4.
Empty the device and make sure that there are no
residues in the measuring chamber.
»
Clean the measuring chamber if necessary (see
Cleaning the
i
Information - Measuring accuracy
Benefit from this rapid and non-destructive measuring procedure and take
several moisture readings of the same sample material.
i
Information - Incorrect readings
Always make sure to select the correct product type and the correct filling
quantity for the material you are measuring. This prevents taking incorrect read-
ings (see
"10.
Faults").
types").
"6. Product
types").
device").
13
"5.3 Hold
values").
"9.3
12
14
15
Page 17