humimeter FS1 & FS1.1 Operating Manual
5.
The measuring process
5.1
Preparing a measurement
To do so:
The device has to have nearly the same temperature than the product being
measured. It is recommended to let your humimeter device adjust to the surrounding
temperature of the material being measured for at least 30 minutes.
1.
Place the empty and clean measuring cup (0.25 liter)
on the switched-off scale (figure 9). Then switch on the
scale.
»
The scale must display 0.0 g with the empty measuring
cup on it. The measuring cup must not be weighed.
2.
Check whether the measuring chamber of the device is
empty. When the device is switched on, there must not
be any material in it.
»
Empty the instrument and clean the measur-
ing chamber if necessary (see
device").
3.
Switch on the device (see
on").
4.
Perform the automatic calibration (see
matic
calibration").
5.
Select the required product type (see
types") by pressing the
Selecting the product
Page 16
"9.3 Cleaning the
"4.1 Switching the device
"4.2 Auto-
"6. Product
or
button (see
type").
9
10
11
"4.4