Troubleshooting
Error message or
warning
Dräger VarioGard
Electronic fault
Error Flash !
Dräger VarioGard
Electronic fault
Error EEPROM !
Dräger VarioGard
Electronic fault
Error RAM I2C !
Dräger VarioGard
Clock setting lost due to
voltage loss.
Set the clock !
Dräger VarioGard
Battery in central unit
is low due to frequent
Alkaline battery low !
voltage losses.
Dräger VarioGard
Battery in central unit is
discharged due to
Alkaline batt. dead !
frequent voltage losses.
38
Cause
Remedy
Must be repaired by
Dräger Service.
Must be repaired by
Dräger Service.
Must be repaired by
Dräger Service.
Set the clock.
As soon as possible:
replace battery.
Replace battery.
Error message or
Cause
warning
Dräger VarioGard
If supply from DC mains:
voltage too low.
Voltage is too low !
If 115 V mains:
power supply set for
230 V.
Electronic fault
Transmitter
Slave (e. g. transmitter)
not connected.
Comms. interrupted !
Slave faulty.
Transmitter
1
Transmitter is
warming up.
Warming up !
Remedy
Check the DC mains.
Convert power supply
to 115 V.
Must be repaired by
Dräger Service.
Check installation and
correct.
Replace slave or
have it repaired by
Dräger Service.
Fault will resolve itself at
the end of the warming-
up time.