13 Specification

Mounting on flow sensor
Integral with sensor
Power supply
Battery
AC mains
Battery power
2 lithium thionyl chloride 'D' batteries.
Note . Recommended/preferred batteries are:
• SAFT LS33600
• Eve ER34615
• GB Cell ER34615
• cT-energy ER34615
The above batteries comply with the safety requirements of
IEC60086-4, and have a maximum peak discharge current of
less than 500mA.
Nominal battery life
Sensor style
Reduced bore
DN40 to 80
DN100 to 300
DN350 to 600
Virtual bore
DN40 to 200
Full bore
DN250 to 600
DN700 to 2400
Probe
300 to 1000 mm
Test conditions:
• acquisition = 15 seconds
• pulse output = 2 Hz @ 2 ms
• alarm output on = 25 %
• logger rate = 1 minute
• no pressure
• integral verification self check = 15 minutes
• ambient temperature = 20 °C
Battery capacity and life are significantly shortened:
• when the operating environment temperature ranges between
–20 and 0 °C or 50 and 70 °C (–4° and 32 °F or 122 and 158 °F)
• when data acquisition is less than 15 seconds
• when pulse output width > 2 ms and output frequency is set high
• with extended NFC interface use
Remote with up to
200 m (650 ft.) cable
Solar
Wind
• OmniCel ER34615
• GEBC ER34615
• LiYa ER34615
• Fanso ER34615H
Size
Integral-mount
Remote-mount
transmitter
transmitter
10 years
7 years
4 years
10 years
4 years
2 years
10 years
Mains supply (option – remote only)
• 85 to 240 V AC, 50/60 Hz @ <3 VA
• Mains connection cable: approx. 3 m (9.8 ft.)
Internal backup power time for:
• DN40 to 200:
• DN250 to 600: 6 days
• DN700 to 2400: 3 days
Renewable power (option)
• Solar or wind
• Input voltage:
• Max. current: 200 mA
Internal backup time for:
• DN40 to 200:
• DN250 to 600: 18 days
• DN700 to 2400: 9 days
Note . Renewable energy generators do not operate at
maximum capacity. For example, low wind speeds, dirty solar
panels and shorter daylight periods all reduce capacity.
Consequently, some installations may require generators with
a capacity greater than the specified 5 W minimum.
Inputs
• IP68 connectors, mains cable (remote only)
• Sensor cable (remote only). ABB sensor cable supplied as
standard. SWA cable available (via adapter box) on
application
• Pressure transducer connection (option)
Configuration/User interface
ABB Velox phone/tablet productivity app. Operating systems
of smartphone or tablet are Android (Kit Kat or newer) and
Windows (10 mobile or newer). Functions of Velox phone/
tablet productivity app include configuration, diagnostics,
logger data retrieval and transmitter flash update.
7 years
5 years
3 years
7 years
3 years
2 years
7 years
16 days
6 to32 V DC @ <5 W
48 days