11. Technical specifications

Model
IH 58 Kids
Type
IH 58
Dimensions
(W x H x D)
110 x 62 x 47 mm
Weight
176 g
Operating
pressure
Approx. 0.25 to 0.5 bar
Atomiser filling
Min. 2 ml
volume
Max. 6 ml
Medication flow
rate
Approx. 0.25 ml/min
Sound pressure
Max. 45 dBA
(acc. to DIN EN 13544-1 section 26)
Mains connection Input: 100 – 240 V~; 50 – 60 Hz; 0.5 A
Output: 5 V
Expected service
400 h
life
Operating
Temperature: +10 °C to +40 °C
conditions
Relative humidity: 10% to 95%
Ambient pressure: 700 hPa to 1060 hPa
Storage and
Temperature: -20 °C to +60 °C
transportation
Relative humidity: 10% to 95%
conditions
Ambient pressure: 700 hPa to 1060 hPa
Aerosol
1) Aerosol delivery: 0.15 ml
properties
2) Aerosol delivery rate: 0.03 ml/min
3) Particle size (MMAD): 4.12 µm
The serial number is located on the device or in the bat-
tery compartment.
Subject to technical changes.
Particle size diagram
Measurements were performed using a sodium fluoride
; 2 A
solution with a "Next Generation Impactor" (NGI).
The diagram may therefore not be applicable to sus-
pensions or highly viscous medications. You can obtain
more detailed information from the manufacturer of your
medication.
12. Replacement parts and wearing
25
parts
Designation
IH 58 Kids year pack contains:
Mouthpiece
Silicone children's mask
Silicone baby mask
Atomiser
Compressed air hose
Filter
Standard IH 58 year pack
includes:
Mouthpiece
Nosepiece with comfort
attachment
Adult mask
Children's mask
Atomiser
Compressed air hose
Filter
Baby mask
Material
REF
602.18
PP
Silicone
Silicone
PP/PC
PVC
PU
602.15
PP
PP/silicone
PVC/aluminium
PVC/aluminium
PP/PC
PVC
PU
PVC
601.31