Interfaces

USB interface

The USB interface type B is used for data exchange
with a PC.
When you connect the instrument to a PC via a standard
USB cable, the SD memory card inserted into the instru-
ment is added to the list of active drives on the PC.
After this, you can carry out all normal file operations on
the SD memory card, e.g. copying and deleting of files.
To learn more about handling of the SD memory card,
see Chapter 3.4 Inserting the SD memory card,
page 3-10.
N o t e
If the instrument is connected to a PC via the
USB interface, it functions as an external
drive. It is then not possible to use the instru-
ment. After disconnecting the USB cable, the
normal operation is possible again.
USM 36

Service interface (LEMO-1B)

The service interface is meant for using the alarm output
and for service purposes for the GE Customer Support.
Pin assignment of the connecting cable
Pin #
Color of wire
1
Brown
2
Red
3
Orange
4
Yellow
5
Green
6
Blue
7
Purple
The alarm signal is available at the pins 3 and 7.
Issue 2 (12/2013)
8 Interfaces and Peripherals
Signal
+5 V
SAP
Alarm
RS232 CTS
RS232 TX
RS232 RX
GND