Standards and reference blocks

CODE NAME
BLOCK
11345A
RB
11345B
RB
11345C
RB
4730
CSK IIA
4730
CSK IIA
4730
CSK IIIA
4730
CSK IIIA
4730
CSK IIIA
4730
CSK IVA
CUSTOM
CUSTOM
5-112
Distance-amplitude curve according to JB/T4730 and GB 11345 (CNDAC)
Ø side-drilled hole (mm) Wall thickness test object (mm)RL (dB)
3
3
3
2
2
1
1
1
Issue 2 (12/2013)
DAC
DAC – 4
DAC – 2
8 ... 46
DAC – 4
46 ... 120
DAC + 2
8 ... 15
DAC + 2
15 ... 46
DAC + 5
46 ... 120
DAC + 10
DAC
DAC
SL (dB)
EL (dB)
DAC – 10
DAC – 16
DAC – 10
DAC – 16
DAC – 8
DAC – 14
DAC – 12
DAC – 18
DAC – 8
DAC – 14
DAC – 6
DAC – 12
DAC – 3
DAC – 9
DAC
DAC – 6
DAC – 10
DAC – 16
DAC
DAC