5 Operation

LARGE (alarm signal)

You can display an alarm signal in the form of a virtual
LED in the far right box in the measurement line on top
of the A-scan. When an alarm is triggered, the color of
the alarm signal changes from green to red.
N o t e
For the configuration of the alarm output, see
Section Configuring the alarm output,
page 5-76.
5-64
– Switch to the second operating level.
– Select the function group EVAL.
– Select the function LARGE or READING 4 and press
one of the selector keys to select the setting VIRTU-
AL LED.
This turns the alarm signal on, and it is displayed next to
the readings on top of the A-scan.
N o t e
You can turn the alarm trigger on by means
of the gate logic (see Section Gate logic,
page 5-74).
All boxes which cannot be displayed are
DISABLED.