GE USM 36 Series Technical Reference And Operating Manual page 164

Ultrasonic flaw detector
Table of Contents

Advertisement

5 Operation
Yearly calibration reminder
– Switch to the second operating level.
– Select the function group CONFIG3.
– Select the function DATE and use the arrow keys to
set the date for the yearly calibration.
– Switch to the function CAL REMINDER and select
the required setup time for the reminder.
– Switch to the function CAL RESET and press one of
the selector keys to acknowledge the reminder.
– Press one of the selector keys for 3 seconds to con-
firm the action. The reminder will then become effec-
tive only after the next calibration date.
5-88
Issue 2 (12/2013)
General setup
USM 36