Preparation For Use; Waveform Selection - Hologic CYNOSURE ELLMAN SURGITRON 4.0 Dual RF/90 IEC Instruction Manual

Table of Contents

Advertisement

4.3.8
4.3.9
4.3.10
4.3.11
4.3.12
Serial Number Label with Fuse and Supply Ratings
5.0

PREPARATION FOR USE

5.0.1
On the Surgitron
position.
5.0.2
Connect the IEC power cord to a 3-wire ground AC power receptacle.
5.0.3
Insert the Neutral Plate male plug into the neutral plate receptacle on the front panel
marked Neutral.
The Neutral Plate should be reliably attached with its entire area attached to the patient's
5.0.4
body as close as possible to the operating field, as per the Safety Instructions specified in
Section 2.1.4.
5.0.5
Select the desired electrode and handpiece and connect them as instructed by their
accompanying instructions for use.
5.0.6
Plug the handpiece into the appropriate receptacle on front panel.
5.0.7
Make sure that the electrode is seated fully so that no conductive shaft is exposed.
5.1

WAVEFORM SELECTION

Reference Section 8 – Power output characteristics for more details about waveforms.
5.1.1
CUT -
Non-Ionization radiation
Equipotential Symbol
Refer to instruction manual/booklet
Keep dry
V. PREPARATION FOR USE
®
Dual/EMC check to make sure that the power switch is in the OFF
For micro-smooth cutting:
Select CUT by pushing SELECT button on the Cutting Mode to turn on the CUT
indicator.
Push Power Up/Down Buttons on the Cutting Mode to select the desired power
level. Depress the CUT Footswitch pedal or Fingerswitch CUT button to activate
the output power.
12

Advertisement

Table of Contents
loading

This manual is also suitable for:

Iec3a-s30-90/110Iec3a-s30-90/220

Table of Contents