:TRIGger:LIN:SIGNal:BAUDrate

Command Syntax
:TRIGger:LIN:SIGNal:BAUDrate <baudrate>
<baudrate> ::= integer from 2400 to 625000 in 100 b/s increments
The :TRIGger:LIN:SIGNal:BAUDrate command sets the standard baud rate
of the LIN signal from 2400 b/s to 625 kb/s in 100 b/s increments. If you
enter a baud rate that is not divisible by 100 b/s, the baud rate is set to
the nearest baud rate divisible by 100 b/s.
Query Syntax
:TRIGger:LIN:SIGNal:BAUDrate?
The :TRIGger:LIN:SIGNal:BAUDrate? query returns the current LIN baud
rate setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 2400 to 625000 in 100 b/s increments