Introduction
®
The LogTag
TRED30-16R temperature recorder features a data
logging memory storing up to 15,905 temperature readings and a
separate statistical memory, storing maximum and minimum reading
as well as alarm duration for each of the last 30 days.
During recording the display shows the current temperature (of the
most recent reading), the status (within or outside the acceptance
range), an alarm trigger summary of up to the last 30 days (today and
29 days previous) of recording, the current time and battery status.
Alarm events can be triggered when a number of readings are outside
pre-set Alarm thresholds and a "day alarm indicator" appears on the
display.
Logged temperature data can be downloaded via a standard LogTag
Interface to the free companion software LogTag
you can display data in chart, list or summary formats. The software
also allows electronic archiving and exporting or transmitting data in
support of sophisticated data management systems.
Features
The TRED30-16R temperature recorder features the familiar LogTag
case layout.
Figure 1: TRED30-16R features
Case
Mounting lug for secure fastening of recorder to fixtures
l
Gold-plated, high-quality temperature sensor socket
l
Robust polycarbonate case, IP64
l
Durable communications contacts
l
Buttons
button; can be used to start and stop the
START
/CLEAR/STOP
l
unit or to clear an alarm. It is also used to exit the statistics
review.
button; can be used to enter the statistics
REVIEW
/MARK
l
review and to scroll through the statistical data directly on the
display. It is also used to place an inspection mark in the data
listing.
Page 4 of 16
®
®
Analyzer, where
TRED30-16R Product User Guide, Revision 1.3, copyright © 2004-2016,
Display
The display shows 'at a glance' if alarm events have occurred for both
the current day and up to 29 days in the past. Details of any alarm
event can be checked directly on the unit by inspecting the statistics
history on the recorder's display or in more detail by downloading the
logged data.
Sensors
The TRED30-16R will accept any sensor from the ST100 product
range.
Checklist - Required Equipment
®
LogTag
TRED30-16R temperature recorder.
l
A computer running Windows XP SP3 or later.
l
®
A LogTag
interface cradle for configuring and downloading the
l
TRED30-16R.
LogTag
®
Analyzer installed and ready to go.
l
Configuring the TRED30-16R
Before a TRED30-16R recorder can be used, it must be configured
with the parameters required for starting and recording temperature
values. This is done using LogTag
used for downloading and analyzing data.
Start the LogTag
®
Analyzer software.
l
Plug a LogTag
®
interface cradle into a USB port and wait for the
l
drivers to be installed. You can use more than one cradle to
configure multiple devices at the same time, however it is
practical to limit the number of units to about 6, using a powered
USB hub.
Insert the TRED30-16R into the interface with the three contacts
l
towards the back until it clicks in place.
Select LogTag - Configure; LogTag
l
ports and display the configuration options for connected
TRED30-16R's.
®
Analyzer software, which is also
®
Analyzer will scan all USB
®
LogTag
Recorders
. All rights reserved.
Need help?
Do you have a question about the TRED30-16R and is the answer not in the manual?