still flashes and -02
l
at the data from 2 days ago.
Yesterday, no temperature values were recorded below the lower
l
alarm threshold, and no alarm event was generated .
The minimum temperature recorded on the day before yesterday
l
was 7.8°C.
Data from 7 days ago
After pressing the
button for a few times (skipping days -3 to -6)
the maximum statistic from seven days ago is displayed:
The
flashes and -07
DAYS
l
being displayed is from seven days ago.
Seven days ago recordings above the upper threshold were recorded,
l
indicated by the upper alarm threshold marker
The duration was 6 hour 30 minutes, which was shorter than the
l
allowed period, so no alarm event was generated
The maximum temperature recorded seven days ago was 10.5°C.
l
Data from 8 days ago
After pressing the
button twice (skipping the minimum statistics
for day -7) the maximum statistic from eight days ago is displayed:
The
flashes and -08
DAYS
l
being displayed is from eight days ago.
Eight days ago recordings above the upper threshold were recorded,
l
indicated by the upper alarm threshold marker
The duration was 5 hour 30 minutes
l
An alarm event was generated
l
8 by itself would cause an alarm event, however the combined
duration is 10 hours 50 minutes, which is more than the allowed 10
accumulative hours.
The maximum temperature recorded eight days ago was 11.5°C.
l
Tip: Pressing and holding the
previous screen, rather than advancing one screen.
Page 13 of 16
still shows, as we are still looking
DAYS
is shown to indicate that the data
is shown to indicate that the data
. Neither duration from day -7 or -
button will get you back to the
Stopping the TRED30-16R
Automatically
The TRED30-16R automatically stops recording temperature when the
maximum number of readings specified during configuration has been
reached. Your unit can also be set up to stop automatically when it is
downloaded. This option needs to be set up at the factory and cannot
be changed during configuration with LogTag
distributor can supply more information about this option.
Manually
You can configure a TRED30-16R so it can be stopped with the
button. This feature is useful when you take the
START
/CLEAR/STOP
recorder out of a shipment and don't want to falsify the statistics with
readings taken after the shipment completion. The stop function is
enabled in the
Advanced Options
Advanced Configuration Options
will stop the unit:
Press and hold the
START
shown in addition to the
Then the
symbol disappears:
Once the
symbol disappears, release the button within 2 seconds.
The recorder will now stop taking readings.
The recorder will not stop if you...
... release the button before the
... keep holding the button for more than 2 seconds after the
symbol disappears.
TRED30-16R Product User Guide, Revision 1.3, copyright © 2004-2016,
®
Analyzer. Your
dialogue during configuration (see
on page 6). When enabled, following
button.
STOPPED
/CLEAR/STOP
symbol.
symbol disappears
®
LogTag
Recorders
is
. All rights reserved.