LED between the buttons
Another LED is located between the Reset and Teach buttons (item 4); this LED acknowledges when you press a
button:
When the key lock
stays on until you interrupt contact with the button.
A description of the Reset and Teach button functions can be found in the next section.

5.1 Reset and Teach buttons

Deactivating the key lock
If you want to operate the keys on the SmartCheck device, you must first deactivate the key lock. Press the Reset
key and then press the Teach key within 2 seconds. The LED between the keys (item 4) flashes to confirm.
After 2 minutes, the key lock is automatically reactivated. The LED between the keys then goes out again.
RESET button
Which function the Reset button executes will depend on how long you press and hold it:
·
If you press and hold the Reset button for longer than 2 seconds, you will reset the current alarms
·
If you press and hold the Reset button for longer than 10 seconds, you will restart the SmartCheck device.
TEACH button
If you press and hold the Teach button for longer than 5 seconds, the learning mode will be restarted for all
measurement jobs that use learning mode.
Further information on the learning mode can be found in the user documentation for the SmartWeb web
application.
Restoring factory settings
CAUTION
32
32
is deactivated, this LED lights up in green as soon as you press one of the capacitive keys. It
·
Before pressing any buttons, please ensure that the device has started up correctly
(observe the status LED after switching on
·
The key lock on the SmartCheck device is enabled by default.
·
You can also go to Device settings > Key settings in the Schaeffler SmartWeb
software to specify the actions permitted using these keys. In this way, you protect the
<SmartCheck/> device against inadvertent operation. For more information, please see
the handbook for the Schaeffler SmartWeb software.
Measurement data and configurations will be irretrievably deleted!
If you restore the Schaeffler SmartCheck device to its factory settings, you will lose all
measurement data and configurations stored in the device!
Before resetting the device, download the measurement data using the Schaeffler
SmartUtility Light software. The full version of Schaeffler SmartUtility also allows you to
download the configuration of the SmartCheck device.