Set-up

You can use the connections on the Schaeffler SmartCheck device to
·
Set up inputs and outputs
·
Set up a PC/Ethernet connection
·
Connect the power supply
To set up and operate the SmartCheck device, you will need the following connection cables, which are not
included in the standard scope of delivery:
·
Input/output cables
·
Ethernet connection cable (Ethernet/PoE)
·
Power supply cable
·
·
General
Please note the following information when connecting cables:
·
Ensure that the sockets for the plug connections are clean and free from impurities. Dirt or moisture in plug
connections can impair signal quality.
·
When securing cables, ensure that they are not subjected to any mechanical strain. If necessary, fit strain relief
accessories.
·
Observe the minimum permissible bending radii for the cables. These can be found in the data sheets from the
cable manufacturer. A minimum bending radius of 60 mm is recommended for cables from the standard range of
accessories from Schaeffler Monitoring Services GmbH.
·
Attach the cables in such a way that they are laid firmly and cannot be knocked.
·
Do not lay signal cables parallel to high-power lines.
14
16
and various connection configurations
16
and
17
The Schaeffler SmartCheck device may only be prepared for use by persons who are verifiably
qualified to do so in accordance with the relevant regulations.
Ensure that the SmartCheck device is de-energised while set-up operations are being carried
out.
Example showing power supply cable from the standard range of accessories