Getting Started with Step-by-Step Position Sweeping

With step-by-step position sweeping, sweeping is exclusively controlled by the test execution
engine which consumes a list of predefined positions and sequentially performs a
measurement at each point.
Step-by-step position sweeping consists of the following typical process flow:
Move the positioner to the desired position
Set up your DUT
Set up the instrument
Take the measurement
Repeat the setup and measurement for each position
Complete the following actions before getting started with step-by-step position sweeping.
Verify that the hardware connections are configured as specified in
Connections
on page 11.
Remove the laser alignment fixture.
Use the mmWave OTA Test Positioner SFP to verify that your DUT can move in all
directions without collision.
Configure TestStand to include the following search paths, including subdirectories, in
the search directories:
C:\Program Files\National Instruments\LabVIEW 2018\user.lib
\OTA Test
C:\Program Files\National Instruments\LabVIEW 2018\instr.lib
C:\Program Files\National Instruments\LabVIEW 2018\vi.lib
\RFmx
Configure TestStand to call LabVIEW Development System (64-bit).
Complete the following steps to perform a full antenna pattern test step-by-step.
1.
Navigate to
C:\Users\Public\Public Documents\National Instruments
\TestStand 2019 (64-bit)\Examples\OTA Test\Sequences
CSV settings file,
2.
Verify that the individual modules are named correctly in MAX. If the names do not
match, either change the name in MAX to match the file or change the file to match your
MAX settings.
3.
Open the
Step_by_Step_TRP.seq
\Public\Public Documents\National Instruments\TestStand 2019
(64-bit)\Examples\OTA Test\Sequences
22 | ni.com | mmWave OTA Validation Test Reference Architecture
Step_by_Step_TRP.csv
.