Operating The Chamber - National Instruments mmWave User Manual

Ota validation test reference architecture
Table of Contents

Advertisement

Required Instrumentation
PXIe-5831 Vector Signal Transceiver—IF and mmWave instrument configuration with
two mmRH-5582 modules (direct TRX ports only)
PXIe-1092 or PXIe-1095 PXI Chassis
PXI Express embedded controller or MXI controller system that meets the system
requirements specified in the chassis documentation.
AC power cable
Optional Items
A 0.9 N · m (8 lb · in.) calibrated torque wrench, 5/16 in. (NI part number 751120-01).
Required Software
You must install the following software to use the mmWave OTA Validation Test Reference
Architecture. NI recommends installing application software before installing the mmWave
OTA Validation Test Software.
mmWave OTA Validation Test Software
TestStand
LabVIEW
RFmx SpecAn
RFmx NR
Note
Refer to the mmWave OTA Validation Test Software Readme for more
information about compatible software versions.
Required Software for Instrumentation
The following instruments require specific drivers that are not installed with the mmWave
OTA Validation Test Software. Refer to the instrument documentation available at ni.com/
manuals for more information about installing required driver software.
PXIe-5831 Vector Signal Transceiver and mmRH-5582 mmWave Radio Head
PXIe-6361 Multifunction I/O Module
PXIe-1095 Chassis

Operating the Chamber

You must complete the following actions when operating the chamber.
Notice
The leakage current from the EMI AC filter inside the chamber may cause
certain residual current circuit breakers (RCCB) to trip. Use the external isolator
provided with the chamber if you are connecting the chamber to an RCCB.
Power On and Reset the Chamber
Complete either of the following actions after you power on the chamber.
If the key switch is set to Manual, press the Reset button.
If the key switch is set to Auto, close the chamber door, then press the Reset button.
10 | ni.com | mmWave OTA Validation Test Reference Architecture
or
PXIe-1092 Chassis

Advertisement

Table of Contents
loading

Table of Contents