Specification - Ametek Land SYSTEM 4 User Manual

Landmark for profibus
Hide thumbs Also See for Land SYSTEM 4:
Table of Contents

Advertisement

User Guide
2

Specification

Input/output accuracy:
Input/output drift:
Quantisation:
Sampling:
Signal retransmission range:
Signal retransmission current:
Signal update rate:
Current output load inc. cable:
Emissivity setting:
Non-greyness setting:
Averager adjustment range:
Peak Picker
decay range:
threshold:
on delay:
off delay:
reset mode:
Track/Hold:
Track/Hold response time:
Alarm:
Trip level:
Relay:
Profibus DP:
PSU*:
Environment:
Vibration:
Humidity:
Sealing:
Ambient:
EMC:
* LAND DPU recommended
< 0.1% of span
< 0.1% of span/°C
16 bit
10ms
A djustable over thermometer span (min 100°)
0 to 20mA/4 to 20mA
1 0ms
500 ohms
0 .200 to 1.000 in 0.001 steps
0 .800 to 1.250 in 0.001 steps
50ms to 512s (63%) in 15 steps
0.25°/s to 512°/s in 12 steps
1° steps
0 to 10s in 0.1s steps
0 to 10s in 0.1s steps
Hold/reset
C ontrolled by external contact closure
5 to 24V d.c.
1 0ms
H i/Lo
1 ° steps
5 0V d.c. at 0.1A
U p to 12MHz
3 0V to 50V d.c. at 300mA (including
thermometer)
24V to 34V a.c. rms
1 g any axis, 10 to 300 Hz
0 to 99% non condensing
I P20
5 to 50°C/41 to 122°F in specification
E N61326
Landmark for Profibus
Page 3