User Guide
3.3.2 CMD (Command) input connection schedule
The CMD (Command) input connection schedule is given in Fig. 5 and Table 1.
The command signal is inactive when the voltage is >3.0V or open circuit
(i.e. Track). The command signal is active when the voltage is <1.5V or short
circuit (i.e. Hold).
The command input can be controlled via a remote switch, as shown in Fig. 6
or via potential-free contacts as shown in Fig. 7.
It is recommended that a twisted pair connection be used.
3.3.3 Output (Retransmission output) connection schedule
The Output (retransmission output) connection schedule is given in Fig. 4 and
Table 1.
It is recommended that a screened twisted pair cable is used.
3.3.4 Thermometer output connection schedule
The thermometer output connection schedule is given in Fig. 4 and Table 1.
It is recommended that 6-way screened cable is used.
3.3.5 Profibus DP slave address
To set the LMP slave address, use the rotary switches located on the top edge
of the LMP, near to the profibus connector.
3.3.6 Profibus DP slave address
To set the LMP slave address, use the rotary switches located on the top edge
of the LMP, near to the profibus connector.
H
0
0
Cycle the power to ensure that any new address is recognised.
3.3.7 Configurator 10101
This terminal is not normally used. See Appendix B.
L
0
= Address 3
0
= Address 21
Landmark for Profibus
Page 9