Eli 380 Preventive Maintenance Record - Welch Allyn ELI 380 Service Manual

Resting electrocardiograph
Hide thumbs Also See for ELI 380:
Table of Contents

Advertisement

ELI 380 Preventive Maintenance Record

Unit Serial #:
Print device configuration (attach to this report).
Perform Visual Inspection of:
Enclosure/Housing
Contamination
Markings and Labeling
Cabling
Integrity of Mechanical Parts
The ELI 380 may have 1 or 2 batteries installed at time of servicing.
Note Battery 1 Date Code
Note Battery 2 Date Code
(Recommend replacement every 2 years.)
Power Testing
Battery 1 – Installed?
Running on DC power only (disconnect AC power cord)
Battery Voltage
State Of Charge
Charge Cycles
Verify Charging is No
On Current (Displayed as Avg. Current)
(Negative reading when on DC)
Battery Health
Running on DC & AC power (connect AC power cord)
Verify AC power LED indicator illuminates (located above power button). Pass / Fail (Circle
One)
Verify AVG Current value gradually changes from negative to positive (the value will remain
at 0 if the battery is fully charged). Pass / Fail (Circle One)
Verify Charging is Yes (if current is flowing to battery). Pass / Fail (Circle One)
Yes
No (if No, skip Battery 1 section)
V
%
Pass / Fail (Circle One)
%
(YYMMDD) or enter N/A if no battery
(YYMMDD) or enter N/A if no battery
A (max draw = 1.6A)
(Recommend replacement when under 70%.)
PREVENTIVE MAINTENANCE
29