Page 1
Digital Clinical Scale with Wi-Fi and Bluetooth 5.0 Owner’s Manual 3300-0192-0M Rev H icon Clinical Scale Owner’s...
Page 2
3300-0192-0M Rev H icon Clinical Scale Owner’s...
Page 3
INTRODUCTION ® Thank you for purchasing our Detecto icon Digital Clinical Scale. It has been manufactured with quality and reliability and has been tested before leaving our factory to ensure accuracy and dependability for years to come. This manual is provided to guide you through installation and operation of your scale. Please read it thoroughly before attempting to install or operate your scale and keep it available for future reference.
Page 4
PROPER DISPOSAL When this device reaches the end of its useful life, it must be properly disposed of. It must not be disposed of as unsorted municipal waste. Within the European Union, this device should be returned to the distributor from where it was purchased for proper disposal. This is in accordance with EU Directive 2002/96/EC.
SPECIFICATIONS 600 lb x 0.2 lb / 1,000 lb x 0.5 lb Capacity: 300 kg x 0.1 kg / 500 kg x 0.2 kg Weighing/Height Units: Pounds/Inch (lb, in) or Kilograms/Centimeters (kg, cm) Platform Size: 17 in W x 17 in D x 1.5 in H (43 cm W x 43 cm D x 3.8 cm D) Overall Dimensions: 17 in W x 18.5 in D x 88.5 in H...
SITE PREPARATION REQUIREMENTS The icon Digital Clinical Scale is a precision weight indicating instrument. As with any precision instrument, it requires an acceptable environment to operate at peak performance and reliability. This section is provided to assist you in obtaining such an environment. Environmental For indoor use only.
UNPACKING Carefully remove the icon Digital Clinical Scale from the shipping carton and inspect it for any damage that may have taken place during shipment. Keep and use the original carton and packing material for return shipment if it should become necessary. The purchaser is responsible for filing all claims for any damages or loss incurred during transit.
Page 10
6. Slide the column cover down onto the base. Note that you may have to pull the cover out slightly to clear the USB port and the AC power jack. 7. Place the scale base cover on the scale base and press down on both sides of the cover.
Page 11
10. Place the bracket over the sonar tube making sure the tube is completely down in the bracket. 11. Next, insert the cable and connector into the hole on the back of the column. 12. Aligning the large end of the slotted holes in the bracket with the screws in the column, (see Step 10 and Step 11) place the sonar tube on the back of the column and pull down to secure the sonar bracket to the column.
Height Rod – Detecto Sonar Height Rod The Detecto Sonar Height Rod connects to the scale by a snap-in modular connector socket. Insert the modular connector of the Sonar Height Rod into the socket on the back of the scale column it locks in place (a clicking sound will be heard when it is locked in place).
Battery Installation/Replacement To install or remove the batteries, the following steps should be followed: 1. Make sure the AC power adapter is unplugged. 2. Remove the platform cover from the scale base. 3. Referring to Figure No. 1, locate the two (2) rectangular panels in the scale base. 4.
PLACING THE SCALE For indoor use only. Place the scale on a flat, level floor or low-cut carpet away from heating and cooling vents. Make certain the AC Power Adapter (if included) cord is out of the way of normal traffic to avoid a trip hazard.
1. Press key to turn the scale on. 2. The scale display will show the software version for a few seconds, the Detecto logo briefly, and then change to a WARNING screen. This screen is displayed to inform and warn you that once you have made the weighing units/height measurement selection, it is PERMANENT and cannot be changed.
KEYPAD FUNCTIONS DO NOT operate the keys or touchscreen with pointed objects (pencils, pens, etc.). Damage to keys or touchscreen resulting from this practice is NOT covered under warranty. This is the Power key. With the scale off, pressing this key will apply power to the scale and turn on the display.
Page 17
This is the Display Mode key. Press and release this key to cause the display to cycle between display modes: Weight/Height/BMI or Weight only. This is the Menu key. Press and release this key to launch the Menu screen. The first screen that is displayed is the Settings Overview screen where you can see all of the settings at a glance.
ANNUNCIATORS The annunciators are displayed on the Weight screen to show that the scale is in the mode corresponding to the annunciator label or that the status indicated by the label is active. Weight Mode WEIGHT: POUNDS/GROSS Power Status Status CENTER ZERO Messages (AC or ##)
OPERATION ALWAYS assist the patient in stepping on and off the scale platform to ensure they do not fall. NEVER leave a patient unattended while they are on the scale platform. Failure to maintain control of the patient at all times can result in serious injury to the patient and/or you.
Body Mass Index (BMI) Operation – (No ID) (Weight, Measure Height, and Calculate BMI) 1. Press the key to turn scale on. 2. If required, press the key to zero weight screen. 3. Assist patient onto scale. NOTE: In order to obtain an accurate height measurement, instruct patient to look straight ahead and not down at scale display.
Body Mass Index (BMI) Operation ( Using Previously Measured Height) 1. Press the key to turn scale on. 2. If required, press the key to zero weight screen. 3. Press the HEIGHT on display screen. 4. If POUNDS/INCHES was selected for Weighing/Height Units, the display will change to show HEIGHT IN INCHES on screen.
Body Mass Index (BMI) Operation with ID ( Using Previously Measured Height) 1. Press the key to turn scale on. 2. If required, press the key to zero weight screen. 3. Press the key. 4. Display will change to show PATIENT ID screen. 5.
MOTHER/BABY WEIGHING ALWAYS assist the mother alone, and when holding the baby/child in stepping on and off the scale platform to ensure they do not fall. NEVER leave the mother or mother holding the baby/child unattended while they are on the scale platform. Failure to maintain control of the mother or mother holding the baby/child at all times can result in serious injury to the mother or mother holding the baby/child and/or you.
1. Press the key to turn the scale on. 2. The scale will show the Detecto logo briefly and then change to the Weight screen. 3. With the Weight screen displayed, press the Menu key. 4. The display will change to show the Settings Overview 1 screen.
Page 25
To Enter Setup, Cont. SETTINGS OVERVIEW 2 <<< Wi-Fi/BLE Module Rev: 1.0.01 Bluetooth: On This value will not be shown if BLE Adv: Icon Scale F8:F0:05:D8:9D:DD Bluetooth: = Off. Wi-Fi: On, Status: Disconnected Network SSID: Dynex617 DHCP: On These values will not be IP: 192.168.0.133 shown if Wi-Fi = Off.
Page 26
Settings Page 1 SETTINGS PAGE 1 Model: Icon Date: 11/6/2019 Time: 12:24:44 Language: English <<< Exit >>> Model: This will change the default settings and operation of the scale. For example, select “Icon” as the model in order to use the stand-up type scale with the sonar height measurement. Touch this key to open the model selection screen.
Page 27
Settings Page 2 SETTINGS PAGE 2 OIML: No Units: lb, in Filter Mode: 0 Zero Tracking: 10 <<< Exit >>> OIML: Touching this key will toggle the OIML setting to Yes or No. NOTE: This setting should be enabled for all EU models to comply with regulations. Allowable settings for OIML are: Yes or No.
Page 28
Settings Page 3 SETTINGS PAGE 3 Capacity 1: 600.0 Capacity 2: 1000.0 Interval 1: 2 Interval 2: 5 <<< Exit >>> Capacity 1: Touching this key will open the capacity screen. This will allow you to set the capacity of the first range.
Page 29
Settings Page 4 SETTINGS PAGE 4 Decimal 1: 1 Decimal 2: 1 Motion Range: 3 Gravity: 0.000000 <<< Exit >>> Decimal 1: Touching this key will open the decimal entry screen. This will allow you to set the decimal point precision for the first weight range. Using the numeric keys, enter a new setting, and then press the Save key.
Page 30
Settings Page 5 SETTINGS PAGE 5 Calibrate Scale Load Cell Trim Power up Zero: No Auto Locking: 5 s <<< Exit >>> Calibrate Scale Touching this key will allow for calibration of the scale. Refer to the CALIBRATION section of this manual for instructions to select the calibration mode and how to perform calibration. Load Cell Trim Touching this key will allow for trimming (adjusting the output of) the load cells in the scale.
Page 31
Press the <<< key to return to the Settings Page 5. Press the Exit key to exit Setup and return to the Weight screen. Requires device to have an activated license. Must purchase Detecto WACONNECT. 3300-0192-0M Rev H icon Clinical Scale Owner’s...
Page 32
Settings Page 7 SETTINGS PAGE 7 Bluetooth: On Wi-Fi: Off SSID: Dynex617 The SSID: value will not be shown if Wi-Fi: = Off. <<< Exit >>> Bluetooth: Touch this key to open a selection screen to turn the Bluetooth 5.0 wireless transmitter On or Off.
Page 33
Settings Page 8 SETTINGS PAGE 8 DHCP: On IP: 192.168.0.133 NM: 255.255.255.0 The settings for NM: and GW: will not be GW: 192.168.0.1 shown if DHCP: = On. <<< Exit DHCP: Touching this key will open an entry screen to allow you to select the DHCP setting. This setting selects whether the Wi-Fi wireless transmitter will use a dynamic IP address assigned by a DHCP server or a static IP address.
1. Press the key to turn the scale on. 2. The scale will show the Detecto logo briefly and then change to the Weight screen. 3. With the Weight screen displayed, press the Menu key. 4. The display will change to show the Settings Overview 1 screen.
Page 35
To Calibrate the Sonar Height Rod 1. Press the 1 key and then press the Save key. 2. The display will show “Clear” for approximately four seconds indicating you should step away from the sensor at a distance of at least 6 feet as it prepares to calibrate. NOTE: Movable objects that might interfere with the sound waves of the sensor should be removed from a 6-foot radius.
Page 36
To Enter A New Height The height in centimeters from the scale platform to the bottom of the sonar height rod sensor has been pre-configured at the factory and should not require changes. If an additional mat or other covering has been added to the scale platform, the new distance, in centimeters from the scale platform to the bottom of the sonar sensor must be measured and entered as accurately as possible to guarantee accurate height measurements from the sonar height rod.
SCALE CALIBRATION Calibration Modes The icon Digital Clinical Scale has four modes that can be used to perform calibration. The modes are as follows: 1 = SmartCal SmartCal is the most precise method of calibration. It requires that a calibrated load be placed over each load cell of the scale platform only once.
Page 38
To Enter Calibration 1. With the scale turned on and the Weight screen displayed, press the Menu key. 2. The display will change to show the Settings Overview 1 screen. SETTINGS OVERVIEW 1 >>> Revision 1.00.09, Display Rev. 1.0.18 Audit Trail -- Cal=4, Cfg=2 Range 1: 600.0 x 0.2 -- 2: 1000.0 x 0.5 Weight Units=lb Height Units=in...
Page 39
To Enter Calibration, Cont. SETTINGS PAGE 5 Calibrate Scale Load Cell Trim Power up Zero: No Auto Locking: 5 s <<< Exit >>> 8. Select Calibrate Scale from the Settings Page 5 screen. 9. The display will change to show the “Number of Points” screen. CALIBRATE SCALE Number of points at which the...
Page 40
1 = SmartCal 1. With the “Number of Points” screen displayed, press the 1 key on the touchscreen and then press the Save key. CALIBRATE SCALE Number of points at which the scale will be calibrated. 1 = SmartCal SAVE CANCEL 2.
Page 41
1 = SmartCal, Cont. 4. The display will change to show “Working” for a few seconds and then change to show “Set test weight over Cell 1”. CALIBRATE SCALE Working To calibrate zero, remove any weight from the scale before proceeding SAVE CANCEL 5.
Page 42
1 = SmartCal, Cont. 8. The display will change to show “Working” for a few seconds and then change to show “Set test weight over Cell 3”. CALIBRATE SCALE Set test weight over Cell 3 and press Save. SAVE CANCEL 9.
Page 43
2 = Two Calibration Points 1. With the “Number of Points” screen displayed, press the 2 key on the touchscreen and then press the Save key. CALIBRATE SCALE Number of points at which the scale will be calibrated. 1 = SmartCal SAVE CANCEL 2.
Page 44
2 = Two Calibration Points, Cont. 5. Place the calibrated test weight (full test load) on the scale, enter the value for the test weight and then press the Save key. CALIBRATE SCALE 1000.0 NOTE: A minimum of This is the known 50% of the scale's test weight capacity is required,...
Page 45
3 = Three Calibration Points 1. With the “Number of Points” screen displayed, press the 3 key on the touchscreen and then press the Save key. CALIBRATE SCALE Number of points at which the scale will be calibrated. 1 = SmartCal SAVE CANCEL 2.
Page 46
3 = Three Calibration Points, Cont. 4. With the “Known test weight” screen displayed, place the first calibrated test weight (half the test load) on the scale, enter the value for the test weight and then press the Save key. CALIBRATE SCALE 500.0 This is the known...
Page 47
4 = Four Calibration Points 1. With the “Number of Points” screen displayed, press the 4 key on the touchscreen and then press the Save key. CALIBRATE SCALE Number of points at which the scale will be calibrated. 1 = SmartCal SAVE CANCEL 2.
Page 48
4 = Four Calibration Points, Cont. 4. With the “Known test weight” screen displayed, place the first calibrated test weight (one fourth the test load) on the scale, enter the setting for the test weight and then press the Save key. CALIBRATE SCALE 250.0 This is the known...
Page 49
4 = Four Calibration Points, Cont. 8. With the “Known test weight” screen displayed, place the third calibrated test weight (full test load) on the scale, enter the setting for the test weight and then press the Save key. CALIBRATE SCALE 1000.0 NOTE: A minimum of 50% of the scale's capacity is...
LOAD CELL TRIM Performing Load Cell Trim 1. With the scale turned on and the Weight screen displayed, press the Menu key. 2. The display will change to show the Settings Overview 1 screen. SETTINGS OVERVIEW 1 >>> Revision 1.00.09, Display Rev. 1.0.18 Audit Trail -- Cal=54, Cfg=12 Range 1: 600.0 x 0.2 -- 2: 1000.0 x 0.5 Weight Units=lb...
Page 51
Performing Load Cell Trim, Cont. SETTINGS PAGE 5 Calibrate Scale Load Cell Trim Power up Zero: No Auto Locking: 4 sec. <<< Exit >>> 8. Select Load Cell Trim from the Settings Page 5 screen. LOAD CELL TRIM 1: 2.941 mV 2: 3.334 mV 3: 3.148 mV 4: 1.747 mV...
Page 52
Performing Load Cell Trim, Cont. 11. Repeat Step 10 until all corners have the same reading. Cell locations are for reference only! Load cell cable in icon column Icon Display 12. Press the Exit key to exit Load Cell Trim and return to the Settings Page 5 screen. 13.
WI-FI AND BLUETOOTH 5.0 The Detecto icon® Digital Clinical Scale has a Wi-Fi and Bluetooth 5.0 wireless transmitter inside the weight display that has been designed for an environment where interfacing the icon scale to a Wi-Fi network or pairing to a Bluetooth-capable device is desired.
Page 54
Enable and Configure Wi-Fi Networking, Cont. 6. The display will change to show the Settings Page 1 screen. SETTINGS PAGE 1 Model: Icon Date: 11/6/2019 Time: 12:24:44 Language: English <<< Exit >>> 7. Press the >>> key to advance to the Settings Page 7 screen. SETTINGS PAGE 7 Bluetooth: Off...
Page 55
Enable and Configure Wi-Fi Networking, Cont. 10. Press the >>> key to advance to the Settings Page 8. SETTINGS PAGE 8 DHCP: On IP: 192.168.0.133 NM: 255.255.255.0 The settings for NM: and GW: will not be GW: 192.168.0.1 shown if DHCP = On. <<<...
Configure Wi-Fi Network Module Without Wi-Fi credentials, the Wi-Fi module will enter listening mode, in which it functions as a Wi-Fi access point. The SSID of the AP is of the form “REDBIRD_WIFI_XX:XX”, where XX:XX is the last 4-digits of the MAC address. Use a Wi-Fi device, such as a Smartphone, tablet, or laptop to connect to the icon scale.
Page 57
Configure Wi-Fi Network Module, Cont. Enter the Network Name (SSID), Password, and Device Name (optional) for the network you wish to connect to. Note that the Device Name can be anything. Click on the Connect button and the module should try to connect to that network. This may take several seconds. The module will store these network credentials and connect to this network each time the scale is turned on.
Changing the Network Credentials 1. With the scale turned on and the Weight screen displayed, press the Menu key. 2. The display will change to show the Settings Overview 1 screen. 3. Press Setup on the touchscreen. 4. The display will change to show the ENTER PASSCODE screen. 5.
WIFI OPERATION Web Server While the scale is connected to a network you can access the hosted web page. To view the web page, proceed to the IP address of the Redbird module with “/webserver.html” added to it. For example, if the IP address is “192.168.0.5”, enter 192.168.0.5/webserver.html and a page like the following should appear.
TCP/IP Connection Continuous output is available by TCP/IP connection to the scale’s IP address at port 10001. The scale will transmit a single set of weight data each time the SMA weight request <LF>W<CR> is received. Each time the SMA weight request <LF>R<CR> is received, the scale will transmit weight data continually until another SMA command is received.
Weight Measurement o Weight Scale Feature Custom Services o Detecto SMA Service Enable Bluetooth (BLE) Networking NOTE: Default Network Settings are: Wi-Fi = Off, Bluetooth = Off, and DHCP = On. 1. With the scale turned on and the Weight screen displayed, press the Menu key.
Page 62
Enable Bluetooth (BLE) Networking, Cont. 4. The display will change to show the ENTER PASSCODE screen. ENTER PASSCODE Enter passcode to access settings. Press CANCEL to exit. SAVE CANCEL 5. Using the 10-key on the touchscreen enter 64870, and then press the Save key. 6.
Enable Bluetooth (BLE) Networking, Cont. IMPORTANT! If an attempt is made to turn Bluetooth (BLE) off when a device is connected to the scale, the display will show the error message Error - BLE Off. SETTINGS PAGE 7 Error – BLE Off Wi-Fi: On SSID: Dynex617 <<<...
Standard Services per Bluetooth SIG Reference adopted specifications at https://www.bluetooth.com/specifications/gatt Device Information Service (0x180A) Characteristics Number Value(s) Attributes Manufacturer 0x2A29 “Detecto” READ Name String Model Number 0x2A24 “Icon” READ String Software Revision 0x2A28 “1.0.XX” software of scale...
BLUETOOTH INTERFACE STANDARD PROTOCOLS Bluetooth Characteristic – Weight_Measurement: 0x2A9D Widely accepted BLE GATT profiles are used to transmit data to other devices/software that have implemented these profiles. Data is passed via BLE using GATT characteristic “Weight Measurement” (0x2A9D) as defined by Bluetooth SIG. Refer to the data table below about Weight Measurement Characteristics.
SMA Commands The SMA commands shown in the WI-FI AND BLUETOOTH SMA COMMANDS table on pages 61 and 62 of this manual can be sent using the Detecto SMA Service’s Rx Cmd characteristic (0x907a0001-8699-47dd-ab30-d7aad5f83e54). Responses will put out a notification on the Tx Cmd characteristic (0x907a0002-8699-47dd- ab30-d7aad5f83e54).
Page 68
WI-FI AND BLUETOOTH SMA COMMANDS, CONT. Command Hex Rep. Response N – scale 0A4E0D Each time sent you will get the next line of scale information information until there is no longer any information. scroll 1. 0A 54 59 50 3A 53 0D = TYP:S 2.
1. Press the key to turn the scale on. 2. The scale will show the Detecto logo briefly and then change to the Weight screen. 3. With the Weight screen displayed, press the Menu key. 4. The display will change to show the Settings Overview 1 screen.
ERROR AND OPERATION MESSAGES The icon scale is equipped with diagnostic software that tests various portions of the scale's circuitry and verifies proper operation. Should a problem be detected; an error or status message will be displayed. The following lists these messages and their meaning. AUTO LOCKED This message appears if the weight/height has been automatically locked.
CARE AND CLEANING NOTE: The scale contains no user-serviceable parts and maintenance should be limited to an occasional cleaning and battery replacement as required. Do not submerge the scale in water, pour or spray water directly on it to clean. The scale is not waterproof and covering it with water will damage it and void the warranty.
PARTS IDENTIFICATION Final Assembly Item Qty. Part Number Description 3300-0140-0A SUB-ASSY, SONAR HR 3300-0141-0A SUB-ASSY, ICON COLUMN 3300-0144-0A SUB ASSY, ICON BASE SONAR BRACKET 3300-0265-08 SCW FLAT-HEAD THREAD CUTTING TYPE 25, 6021-1293 #4-24 X 1/4, PHIL. DR. Z-PLATE SCW ROUND-HEAD BLUNT SS. SHEET METAL #6 X .25 6021-6017 3300-0151-0A SUB-ASSY, DISPLAY, ICON...
PARTS IDENTIFICATION Sonar HR Sub-Assembly Item Qty. Part Number Description 3300-0022-08 HEIGHT SENSOR TUBE 3300-0060-08 SONAR FRONT ENCLOSURE 3300-0061-08 SONAR REAR ENCLOSURE 3300-0135-0A CABLE, ICON/APEX SONAR SCW FLAT-HEAD THREAD CUTTING TYPE 25, 6021-1293 #4-24 X 1/4, PHIL. DR. Z-PLATE SCW PAN HEAD PHILLIPS #1-32X3/8” THREAD 6021-2078 FORMING, 18-8 SS 3300-0039-2A...
PARTS IDENTIFICATION Column Sub-Assembly Item Qty. Part Number Description 3300-0067-08 THROAT REAR 3300-0068-08 THROAT FRONT 3300-0087-08 COLUMN, DRILLED & SONAR 3300-0278-1A CABLE, APEX COLUMN, SONAR SCW FLAT-HEAD THREAD CUTTING TYPE 25, 6021-1293 #4-24 X 1/4, PHIL. DR. Z-PLATE 6021-1509 SCW PAN-HEAD, SHEET METAL #8X.75 6650-1114 O-RING 2 ID X 2-3/16 OD X 3/32 THK 6710-1021...
(2) years from date of shipment. Detecto shall be the sole judge of what constitutes a defect.
Page 79
This warranty sets forth the extent of our liability for breach of any warranty or deficiency in connection with the sale or use of the product. Detecto will not be liable for consequential damages of any nature, including but not limited to, loss of profit, delays or expenses, whether based on tort or contract.
Page 80
Cardinal Scale Manufacturing Co. 102 E. Daugherty, Webb City, MO 64870 USA Ph: 417-673-4631 or 1-800-641-2008 Fax: 417-673-2153 Technical Support: 1-866-254-8261 Printed in USA E-mail: tech@cardet.com 3300-0192-0M Rev H 03/20 3300-0192-0M Rev H icon Clinical Scale Owner’s...
Need help?
Do you have a question about the ICON Series and is the answer not in the manual?
Questions and answers