Table 2
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge IEC61000-4-5
Voltage dips,short interruptions and voltage
variations on power supply input lines
IEC 61000-4-11
Magnetic field
IEC 61000-4-8
Conduced RF
IEC61000-4-6
Radiated RF IEC61000-4-3
56
English
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601 test level
±8 kV contact, ±2kV, ±4 kV, ±8kV, ±15 kV air
not applicable
not applicable
not applicable
30 A/m
50 Hz/60 Hz
not applicable
10 V/m
80 MHz - 2,7 GHz
80% AM op 1kHz
NOTE UT is the a.c. mains voltage prior to application of the test level.
Compliance level
±8 kV contact, ±2kV, ±4 kV, ±8kV, ±15 kV air
not applicable
not applicable
not applicable
30 A/m
50 Hz/60 Hz
not applicable
10 V/m
80 MHz - 2,7 GHz
80% AM op 1kHz