10. Troubleshooting (2)
PrOBLEM
LCD Display shows
"EE"
LCD Display shows
abnormal result
no response when you
press button or load
battery
z
MAInTenAnce
z
1.
Do not drop this monitor or subject it to strong impact.
2.
avoid high temperature and solarization. Do not immerse the monitor
Σ
IVD
1
in water as this will result in damage to the monitor.
z
Σ
IVD
3. If this monitor is stored near freezing, allow it to acclimate to room
1
temperature before use.
PHTHALATES
4.
Do not attempt to disassemble this monitor.
PHTHALATES
5. If you do not use the monitor for a long time, please remove the batteries.
Σ
IVD
1
6. It is recommended the performance should be checked every 2 years
or after repair. please contact the service center.
7. Clean the monitor with a dry, soft cloth or a soft cloth squeezed well after
PHTHALATES
moistened with water, diluted disinfectant alcohol, or diluted detergent.
8. no component can be maintained by user in the monitor. the circuit
diagrams, component part lists, descriptions, calibration instructions,
or other information which will assist the user's appropriately qualified
technical personnel to repair those parts of equipment which are
designated repairably can be supplied.
9. the monitor can maintain the safety and performance characteristics
for a minimum of 10,000 measurements or three years, and the cuff can
maintain the performance characteristics for a minimum of 1000
measurements.
POSSIBLE CAUSE
the cuff was not properly
applied or the rubber tube
was bent or pressed
the cuff position was not
correct or it was not
properly tightened
Body posture was not
correct during testing
Speaking, arm or body
movement, angry, excited
or nervous during testing
Incorrect operation, or strong
electromagnetic interference
ENGLISH
SOLUtION
review the cuff applying
and testing sections of the
instructions and re-test
apply the cuff correctly
and try again
review the body posture
and testing sections of the
instructions and re-test
re-test when calm and
without speaking or moving
during the test
take out batteries for five
minutes, and then reinstall
all batteries