Section 2
* These functions are used by service technicians to troubleshoot the instrument.
Page 36
Instrument Menu Structure
CAL DEFAULTS—(Calibration Default Settings) Press
the calibration settings back to factory defaults and set the Gain to 1.0
and the Offset to 0. Use this feature to remove calibrations that are
causing a MARG G or MARG Z to be displayed.
OFFSET—Press
ENTER
correction factor. The factory default is 0.
GAIN—Press
ENTER
factor. The factory default for gain is 1, however, readings of 0.8 to 1.25
are acceptable.
DISPLAY TEST—Press
to visually check that the complete display is functioning.
RECMIN—Recorder signal test. Press
output to its minimum (normally 4 mA) and hold it there until the next
analysis is complete.
RECMAX—Recorder signal test. Press
output to its maximum (normally 20 mA) and will hold it there until the next
analysis is complete.
Note: To hold the signal longer, press the
scroll to
MAINTENANCE
and press
again.
ENTER
REC ADJUST—Press
Recorder Maximum values. Use this feature to offset the 4 or 20 mA
output signal to force a chart recorder to read zero or full scale in case of
a small discrepancy. Use the
or
then use the
RECMIN
RECMAX
the value. Press
ENTER
return to
.
SETUP
Note: The numbers on the display are A/D counts and cannot be used to
determine the recorder setting.
RELAY TEST—Press
sequentially turn on ALARM1 and ALARM 2. Automatically clears both
alarms to complete the test.
REF*—Momentarily displays the last Reference A/D counts.
SAMPLE*—Momentarily displays the last Sample A/D counts.
ZERO*—Momentarily displays the last Zero A/D counts.
CYCLE TIME*—Momentarily displays the Motor On time (in seconds).
POWER FAIL WARNING—Sets the power fail warning to ON or OFF.
Press
then use the
ENTER
Press
to select. Press
ENTER
. If the power fail warning is turned on and the power is interrupted,
SETUP
a system warning will be generated when power is restored. The factory
default is off.
to momentarily display the current Zero
to momentarily display the current Gain correction
to momentarily display all LCD segments
ENTER
to drive the Recorder
ENTER
to drive the Recorder
ENTER
key then the
MENU
. Press
then scroll down to
ENTER
to modify the Recorder Minimum and
ENTER
and
RIGHT
LEFT ARROW
and
UP
DOWN ARROW
to select. Press
to accept changes and
EXIT
to momentarily turn the relays off, then
ENTER
keys to toggle between ON and OFF.
ARROW
to accept changes and return to
EXIT
to return
ENTER
key to
DOWN ARROW
HOLD OUTPUTS
keys to select
keys to adjust