PROFESSIONAL MEDICAL PRODUCTS
PALLONI AUTOCLAVABILI IN SILICONE
SILICONE AUTOCLAVABLE RESUSCITATORS
INSUFFLATEURS EN SILICONE AUTOCLAVABLES
BOLSAS AUTOCLAVABLES DE SILICONA
BALÕES DE SILICONE AUTOCLAVÁVEIS
Manuale d'uso - User manual
Manuel de l'utilisateur - Guía de uso
Guia para utilização
ATTENZIONE: Gli operatori devono leggere e capire
completamente questo manuale prima di utilizzare il prodotto.
ATTENTION: The operators must carefully read and completely
understand the present manual before using the product.
AVIS: Les opérateurs doivent lire et bien
comprendre ce manuel avant d'utiliser le produit.
ATENCIÓN: Los operadores tienen que leer y entender completamente
este manual antes de utilizar el producto.
ATENÇÃO: Os operadores devem ler e entender
completamente este manual antes de usar o produto.
34244 - 34245 - 34246 - 34247
Gima S.p.A.
Via Marconi, 1
20060 Gessate (MI) Italy
Made in Taiwan
Gima S.p.A.
Via Marconi, 1 - 20060 Gessate (MI) Italy
gima@gimaitaly.com - export@gimaitaly.com
www.gimaitaly.com
0476

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 34244 and is the answer not in the manual?

Questions and answers

Summary of Contents for Gima 34244

  • Page 1 ATENCIÓN: Los operadores tienen que leer y entender completamente este manual antes de utilizar el producto. ATENÇÃO: Os operadores devem ler e entender completamente este manual antes de usar o produto. 34244 - 34245 - 34246 - 34247 0476 Gima S.p.A. Via Marconi, 1...
  • Page 2: Product Description And Intended Use

    The designs are also different according to Adult, Child or Infant by using different compressing frequency, they also come with different sizes to meet all patients’ need for oxygen. Codes 34244, 34246 and 34247 are equipped with a POP-Off valve, the model code 34245 does not include the POP-Off valve in the standard equipment The adult size has a Pressure Relief Valve design (as an option).
  • Page 3: Principles Of Operation

    ENGLISH Principles of operation Assembly view of an artificial respiration system (A) (A) Non-rebreathing (duckbill) Valve (B) Silicone Bag (C) Reservoir Valve (D) Oxygen Reservoir. The Reservoir Valve and the Oxygen Reservoir should be removed if supplemental oxygen is not to be supplied.
  • Page 4 ENGLISH Principle Drawing (D) GIMA Infant and Child Resuscitator are both equipped with Pressure Relief Valves, automatically pro- vide and adjust the pressure in the lung, and keep it within 40 cmH O +/- 5 cmH O, any pressure exceed this standard will cause the Pressure Relief Valve to jump off and push the pressure out to ensure the patient’s safety.
  • Page 5 ENGLISH 34245 – 34246 - 34247...
  • Page 6: Cleaning, Disinfection And Sterilization

    ENGLISH Operating Instruction 1. Place the patient on back, pull his chin upward as possible to keep the airway and the mouth cavity in alliance line, so the patient can breathe smoothly. 2. Clean all visible foreign material inside the mouth and the throat.
  • Page 7 ENGLISH The cleaning-disinfection-sterilization process The following steps are generally recommended. Select proper methods for the resuscitator parts in question according to the table. Cleaning methods • Disassemble the resuscitator following the assembly view and order. Do not disassemble the pres- sure relief valve spring, just rinse it directly.
  • Page 8 ENGLISH Visual inspection through package assembly 1. After cleaning-inspection-sterilization process, let the resuscitator parts stay at room temperature to dry (air dry). Wait until all parts are dry. No drying agent is needed. 2. After cleaning-disinfection-sterilization carefully inspect all parts for damage or excessive wear and replace them if necessary.
  • Page 9 ENGLISH Testing the Resuscitator The GIMA Manual Resuscitator should be tested as follows: • When first using the new Resuscitator • After cleaning and sterilizing • After any new parts have been fitted • Monthly, if the Resuscitator is not frequently used.
  • Page 10: Specification & Performance

    ENGLISH 3. Cycle the Resuscitator through several ventilations. The test lung should inflate during inspiration and deflate during exhalation. Check for leakage at all joints and connections. Ensure that the Resusci- tator refills promptly and properly and that all valves are operating correctly. If not, repeat the tests above to find where the problem lies.
  • Page 11 ENGLISH Technical specifications and performance characteristics of the device Dead Space Pressure Relief Valve nonrebreathing valve 7 mL Child and infant 40±5 cmH adult mask 150 mL Adult 60±10 cmH child mask 95 mL infant mask 28 mL Bag volume Stroke volume Reservoir volume Suitable body weight...
  • Page 12 ENGLISH Stroke Volume Range Using one hand Using two hands Adult 700 ml 900 ml Child 300 ml 350 ml Infant 150 ml 225 ml Oxygen concentration for Adult model Values in brackets are without an oxygen reservoir. Tidal Volume (ml) x Ventilation Rate with Reservoir [without Reservoir] Oxygen Flow (LPM) 600x12 600x20...
  • Page 13 ENGLISH The GIMA Peep Valve (Accessory Series) Terminal Positive Pressure Adjustment Valve (for one patient use) 1. 2-10 cmH O Terminal Positive Pressure Adjustment Valve for one patient use 2. 5-20 cmH O Terminal Positive Pressure Adjustment Valve for one patient use 3.
  • Page 14 The warranty is valid for 12 months from the date of supply of GIMA. During the period of validity of the warranty, GIMA will repair and/or replace free of charge all the defect- ed parts due to production reasons.