Calibration Instructions : KWD 500
MODELS : KWD 500
ACTION : Calibration
_________________________________________________________________________________________________
Calibration method
1. Scale should be powered on for at least 3 minutes before
proceeding to calibration.
2. Locate calibration switch found on underside of scale and move it
to "Cal " position.
3. Power on scale. –CAL will be displayed.
4. Press and hold the ZERO and ON/OFF keys at the same time.
5. Display will read either USC or PC.
6. Press the ZERO key to select USC and then press the TARE key
to confirm selection.
7. The maximum weight capacity value in pounds ( lbs. ) will be
displayed.
8. Press the ZERO key to scroll down the different values. When
correct value needed for that model is displayed ,press the TARE
key to confirm selection.
9. Display will know read " 0 " , if not press the ZERO key.
10. Under conditions of neither wind drafts nor vibrations, load the
required capacity unto platter. When display is stable press the
TARE key to confirm.
11. display will read END for 1 second and then CAL.