Page 1
Kilotech KRS SWP Series Counter-top weighing scale NTEP (USA) and Measurement Canada approved www.kilotech.com...
Page 2
NOTE: Le manuel en Français est disponible au www.kilotech.com. Thank you for purchasing a Kilotech Model KRS SWP Weighing Scale. Please read all operating instructions carefully before using the scale for the first time. This scale is approved by the US and Canadian governments and can be used in a “Legal for Trade”...
Contents 1.0 Precaution before Using the Scale 1.1 Environment 1.2 Protection in the transportation 1.3 Leveling the scale 1.4 Turn on scale 2.0 Explanation of Display Symbols 3.0 Keypad Functions 4.0 Operations 5.0 Scale Parameters Setting 5.1 Backlight type 5.2 Auto shut. Off time 6.0 Power Supply 7.0 Error Message...
1.1 Environment The scale should always be used for its intended use (see www.kilotech.com) and its intended capacity. This scale is not protected against water; DO NOT spray with liquids as this could damage the electronics or could cause serious injury.
1.2 Protection in the transportation Before the initial use, please refer to the following drawing to take off the protection screw. When packing the scale for transportation, install the protection screw on the scale in order to avoid any damage to sensitive components. 1.3 Leveling the Scale The scale is equipped with a level indicator on the left bottom side of the front panel and four adjustable leveling feet.
2.0 Explanation of Display Symbols LCD Display Meaning of the Symbols (Zero): Scale is in ZERO mode. NET: Scale is in TARE mode. : Low Voltage (Stable): Scale is in stable mode. kg/ lb: Indicates the current measuring units in weighing mode.
3.0 Keypad Functions Button Function Zero the scale Zero/Enter Accepts the current setting on the display. Move the settings backward in CAL parameter settings Move the settings forward in CAL parameter settings Select the options in settings Select the options in settings 4.0 Operations Turn on the scale;...
5.0 Scale Parameters Setting Press and ZERO/ENTER key at the same time in weighing or counting mode to enter into parameters setting. 5.1 Backlight type to s et the backlight type to be “bl. 1”,“bl. 2”or “bl. 3”. Press Please refer to the following. Backlight type Description bl.
6.0 Power Supply & Battery Operation POWER SUPPLY DC 12V/1A 7.0 Error message During the initial power-on testing it is possible the scale may show error message. The meaning of the error messages is described below. Error Possible causes Remedy codes EPROM data lose.
This should be done by a specialized scale professional. If your distributor is an industrial scale dealer, he will be able to do this for you. If you do not know who to contact, call the Kilotech customer service desk at 800-694-4445 (Canada) or 877-328-5988 (USA) and we will recommend someone.
Page 11
C Press key to select the capacity to be: 3kg or 6kg. Press ENTER key to confirm and move to “Resolution”. Resolution The display shows as below: 3000 Press key to select the resolution to be: 3000 or 6000. Press ENTER key to confirm and start weight loading.
Page 12
Press ENTER key to confirm the current zero point, and then the display shows the calibration weight value for the first time (1/3 of full capacity) which should be put on. L 1.000 Put on the weight refer to the display.
Page 13
display is stable to confirm. Then the scale enters into parameters setup mode. Note: During the weight loading procedure, press check the internal count (Press again to return). Parameters setup The display shows as below: Press ENTER key to set the “Auto zero tracking range”. Auto zero tracking The display shows as below: ...
Page 14
Press key to choose the range to be off, 0.2e, 04e or 0.6e. (Default: 0.6e) Press ENTER key to confirm and move to “Units selection”. Units selection The display shows as below: U kg Press key to select the unit: kg or lb.
Page 15
Type2: SHARP ER-Axxx, ER-A450T, New SANYO ECRs using RS-232 and others Type3: Most P.O.S. system Type4: CRS, NCR2170, SAMSUNG ER-5100, ER5115 and Many other ECRs Type5: NCI General, SAMSUNG ER5100, Most P.O.S. Software Type6: SAMSUNG ER-670 Type7.1: Continuous output Type 7.2: Output when stable (Once there is a weight change exceeds 1e, it will output) Note: For type0-type5, the Serial communication format is:...
Page 16
Parity bit: None Stop bit: 1 Press ENTER key to confirm. The calibration procedure is finished. The scale will re-start and enter into normal weighing mode. ...
Need help?
Do you have a question about the KRS SWP Series and is the answer not in the manual?
Questions and answers