The amount of thermal radiation emitted from an object de-
pends on the specific radiation properties of the material and its
surface. Non-contact temperature measurement requires that
you determine the material constant (emissivity) of the target
object prior to first-time use
7.7

Automatic switch-off

In the default setting, the CellaPort switches off automatically after 2
minutes if none of the control keys are pressed. Automatic switch-off can
be configured from 1 – 60 minutes. The CellaPort measures continuously
when it is in "automatic disconnection deactivated" mode.
8
Setting parameters at the pyrometer (basic configura-
tion)
Use the buttons ▲▼ and the „MODE" button on the rear panel to access
and configure parameters. With these buttons you can view and adjust
all settings required for operating the pyrometer.
Menu structure:
MOD
E
Normal
operating mode
(temperature)
1. Press the MODE button while in normal operating mode to switch
to "configuration layer" mode.
2. Use ▲▼ to select the configuration layer for the parameter you
wish to set.
3. Press MODE to confirm. Press ▲▼ to select the particular pa-
rameter.
4. Press MODE to confirm. Press ▲▼ to adjust the parameter value.
5. Press MODE again to end. Press ▲▼ to select
Whenever a control key has not been pressed within 30 seconds, the py-
rometer will automatically return to normal operating mode. The modified
value is applied.
MOD
E
Select
config. layer
C000-
C999
▲▼
9
MOD
E
Select
parameter
EPS1
▲▼
END.
Adjust
parameter
99.9
▲▼