Guidance and manufacturer's declaration - electromagnetic emissions
The PRO-110/PRO-115 is intended for use in the electromagnetic environment specified
below. The customer or the user of the PRO-110/PRO-115 should assure that it is used in
such an environment
Immunity Test
Conducted RF
IEC
61000-4-6
Radiated RF
IEC 61000-4-3
20
M_PRO-110_PRO-115_EN_RO_BG_2916.indd 20
3 A/m
3 A/m
IEC 60601
Compliance
level
Test level
3Vrms
3V
150KHz to
80MHz
3 V/m
3 V/m
80MHz to
2.5GHz
PRO-110 / PRO-115
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer
to any part of the PRO-110/PRO-115,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
3,5
d=[
] √P
V
1
3,5
d=[
] √P 80MHz to 800 MHz
E
1
7
d=[
] √P 80MHz to 2,5 GHz
E
1
where p is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m)
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,a should be less than the compliance
level in each frequency range.b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOT
NOT
affec
a: F
b:
Reco
comm
21.07.2016 11:39:03