7. OPERATING

Hold the nebulizer strictly vertically.
WARNING! Do not incline the nebulizer at more than 45 degrees.
If the angle is greater than 45 degree, the drug can flow out to your mouth.
To start the treatment, set the switch into the «I» position. The compressor will start
working and a nebulizing will begin
Do the treatment in accordance with your doctor's instructions.
Inhale the aerosol solution using the prescribed accessory.
WARNING! The temperature of inhaled aerosol depends on the temperature of the
ambient atmosphere and the temperature of the drug solution. If the solution was
stored in refrigerator for a long period of time, it is recommended to raise the
temperature to 16-20С°.
Use of mouthpiece:
Take the mouthpiece to your mouth and breathe smoothly during the inhalation.
Using of mouthpiece is recommended in treatment of lower respiratory tract in grown-
ups and children above 5 years old.
Use of masks:
Put on a mask in the way that it will close your nose and mouth, and conduct inhalation
of the drug. Inhale and exhale through the mask.
Using of mask is recommended in treatment of upper respiratory tract. It allows for
irrigation of the whole nasal and pharynx cavities, and also larynx and trachea.
Child mask is recommended for use for children aged from 1 to 5 years old Infant mask
for children under 1 year old.
When the treatment has been completed, switch the unit off by setting the switch into
the «O» position and disconnect the plug from the wall socket.
Wash the nebulizer and its accessories as described in the section «9. Cleaning, Mainte-
nance and Storage».
Condensate can form in the air tube. In this case detach the air tube from the nebulizer,
turn on the compressor and dry the air tube till the elimination of all the liquid.
WARNING! Do not store the air tube if there is condensate or liquid inside. It can
lead to bacterial infection.
Switch off the device from the socket.
Rinse your mouth with boiled water of the ambient temperature after the procedure. If
you used mask – rinse your eyes and face with water.
WARNING! The device can work for 30 min without a pause, than you should let
the device to cool down for 30 min.