Pulse profiling application
Hide thumbs Also See for PowerSensor+ LB480A:
Table of Contents

Advertisement

Models LB480A/LB680A USB PowerSensor+™
Product Manual PPA (Pulse Profiling Application)

Poles

The Poles drop down allows selection of one, two or four filter poles as shown below. The default value is four poles
with option 004 and one pole without option 004.
The number of Poles has the effect of determining the sharpness of the frequency roll off depending on the filter
selection. The greater the number of Poles the steeper the roll off and again, the reduction in the high frequency
component of the signal. More Poles then decrease roll off but have the affect of increasing both the rise and fall times
of the displayed signal (as time relates inversely to frequency).
NOTE: The Filter and Poles selections are only active with LB480A Option 004. Otherwise the selections will be
inactive and set to 100 kHz and one pole.

Uncalibrated Measurement Message

You will see the popup below if the measurement becomes uncalibrated due to the filter, poles or sweep time settings.
Also, an "Uncal Meas" label will appear on both the Panorama and Measurement view grids. Adjust the settings
accordingly until the error message stops.