Summary of Contents for BOC Edwards Seiko Seiki STPH803C
Page 1
B717-00-880 Issue E Instruction Manual Seiko Seiki STPH803C/H1303C/H803CI (P033) Turbomolecular Pump System (Document number MT-33E-002-F & MT-33E-0A1-C) Manor Royal, Crawley, West Sussex, RH10 2LW, UK Telephone: +44(0)1293 528844 Fax: +44(0)1293 533453 http://www.bocedwards.com...
Page 2
Manual carefully before using the STP pump. Keep this Manual in a place where you can quickly access it at any time. BOC Edwards Japan Limited Dec. 2003 Copyright 2003 BOC Edwards Japan Limited. All rights reserved. Printed in Japan.
Page 3
Safety Precautions as well as the general safety rules (your country’s laws, regulations, safety standards and so on). If the equipment is used in a manner not specified by BOC Edwards, the protection provided by the equipment may be impaired.
Page 4
STP-H803/H1303 Series Instruction Manual The STP pump is provided with a high-speed rotor. Secure the STP ◇ pump according to the specified method. Failure to do so may lead to serious personal injury, product and/or peripheral equipment damage if any abnormality/error occurs in the rotor. The STP pump operates at high temperatures while the baking heater ◇...
Page 5
STP-H803/H1303 Series Instruction Manual NEVER use any gas that is not specified as usable in this Manual. ◇ The use of such gas may corrode the STP pump and damage it. A hazardous live voltage may exist at connector/terminal that marked ◇...
Page 6
(the STP pump serial number, the STP control unit serial number, and the STP connection cable length) is changed. Use the STP connection cable that BOC Edwards has specified. ◇ The use of different cables may result in product damage.
Page 7
STP-H803/H1303 Series Instruction Manual NEVER turn OFF the primary power (DO NOT switch the breaker ◇ "OFF") while the STP pump is rotating. Doing so may result in product damage. Perform investigations into probable causes and remove them before ◇ restarting the STP pump in the event of the occurrence of a problem.
Page 8
STP-H803/H1303 Series Instruction Manual INTRODUCTION Thank you very much for purchasing BOC Edwards’ turbomolecular pump. The turbomolecular pump is designed to be installed in the vacuum equipment to exhaust gases from it. This manual covers all items necessary to ensure safe installation, operation and...
1) No part of this manual may be reproduced in any form by any means without prior written permission from BOC Edwards. 2) BOC Edwards pursues a policy of continuing improvement in design and performance of this product. The right is, therefore, reserved to vary specifications and design without notice.
1) This warranty applies only to the product delivered from BOC Edwards to the customer. 2) If any defect is found during this period, BOC Edwards will, at its option, repair or recondition the product free of charge. The costs for repair or replacement of the product after the warranty period has passed will be at your own charge.
Page 11
STP-H803/H1303 Series Instruction Manual external factors; 13) Deterioration in the basic performance due to the use of the product beyond limits of the use; 14) Any direct, incidental or consequential damage resulting from the use of the product; 15) When continuously operated without overhaul after the WARNING indication ("WARNING"...
STP-H803/H1303 Series Instruction Manual TABLE OF CONTENTS SAFETY PRECAUTION INTRODUCTION APPLIED STANDARDS PRECAUTIONS REQUEST LIMITED WARRANTY 1 Precautions for Safe Operation of the STP Pump..........1-1 Usable Gases ........................1-1 Maintenance and Inspection Precautions................1-1 Labels..........................1-2 2 Operation Principle of the STP Pump ..............2-1 3 Unpacking ......................3-1 Unpacking the STP Pump and the STP Control Unit ............
Page 13
STP-H803/H1303 Series Instruction Manual Attaching the STP Control Unit to a Rack................6-8 Cable Connection ....................... 6-9 6.4.1 Name and Dimensions of Each Cable..............6-9 6.4.2 How to Connect the Cables ................. 6-12 Connecting to Semiconductor Equipment ................ 6-14 6.5.1 Connecting to Power....................
Page 14
STP-H803/H1303 Series Instruction Manual 11.3.1 Abnormalities When Powering ON ..............11-8 11.3.2 Abnormalities When Performing the STP Pump Start Operation......11-8 11.3.3 Other Abnormalities ..................... 11-9 11.4 "WARNING" Message Display Function................. 11-10 11.4.1 First Damage Limit..................... 11-10 11.4.2 Second Damage Limit..................11-11 11.5 "ERROR RECORD"...
Page 15
STP-H803/H1303 Series Instruction Manual TABLES Table 3.1 Accessories ........................3-2 Table 4.1 Tightening torque of bolt....................4-9 Table 4.2 Maximum Torque predicted and Recommended securing bolt for inlet port flange ..4-10 Table 4.3 Number of Claw Clamps by Size of Flange..............4-11 Table 6.1 Connecting the Power Cable..................
Page 16
STP-H803/H1303 Series Instruction Manual FIGURES Figure 2.1 Cross Sectional View of the STP Pump ................. 2-2 Figure 3.1 Example of Lifting the STP Pump................... 3-2 Figure 4.1 Configuration of the STP Pump..................4-2 Figure 4.2 Installation of the STP Pump to the Vacuum Equipment ..........4-6 Figure 4.3 STP Pump Installation Positions ..................
The STP series models are chemical specific pumps: chlorine or fluorine system gases can be used. When you use gases including alkaline metals, but excluding Li, gases including Ga, Hg, In, or Sn, or HBr, contact BOC Edwards. To prevent an accident, confirm the characteristics of gases to be ◇...
STP-H803/H1303 Series Instruction Manual Labels The following labels are affixed to the STP pump and STP control unit. Read the contents of the labels before operation. For the positions of the labels, see Section13, "Label Affixing Position." 1) STP Pump Caution Label This label describes precautions for operating the STP pump.
Page 19
STP-H803/H1303 Series Instruction Manual 4) STP Pump Installation Warning Label This label describes installation of the STP pump. Install the STP pump according to the precautions of Section 4, "Installation of the STP Pump." 警告 WARNING ポンプは取扱説明書の指示 Install pump securely に従い確実に固定してくだ...
Page 20
STP-H803/H1303 Series Instruction Manual 7) Rotational Direction Instruction Label This label describes the rotational direction of the STP pump. The STP pump rotates in this direction. ROTATION 8) High Voltage Device Caution Label The STP control unit is equipped with a high voltage device. This label warns operators to pay attention to the high voltage device.
Page 21
STP-H803/H1303 Series Instruction Manual 10) Hot Surface Warning Label This label instructs operators so as not to touch the hot surface of the STP pump. The use of the TMS unit may lead to a considerable rise in temperatures outside the STP pump. This label warns operators so as not to burn hands.
STP-H803/H1303 Series Instruction Manual Operation Principle of the STP Pump The STP series pump is a series of a magnetically-levitated turbomolecular pumps, featuring the following: • Oil free • Low vibration • High reliability The STP pump is configured so that rotor blade (1) and stator blade (2) are aligned alternately in the axial direction.
• Check the package for bruises, breakage, wetness, and other. If there is any abnormality/error or it is judged necessary to return the product, contact BOC Edwards or the selling agency. • Check the contents of the package. See Section 3.2, "Accessories."...
O-ring washer for purge port KF10 4 legs are attached to the STP pump Rubber foot for leg Instruction Manual : The standard cable length is 5 m. Both 10 m and 20 m are available. Contact BOC Edwards on other specifications.
STP-H803/H1303 Series Instruction Manual Installation of the STP Pump Name and Function of Each Part (1) Inlet Port Flange (VG , ICF , ISO, and other) ・ Connected to the vacuum equipment (at the high vacuum side). (2) Outlet Port Flange (KF ・...
STP-H803/H1303 Series Instruction Manual (1) Inlet port flange (6) Temperature sensor connector (5) Cooling water port (2) Outlet port flange (3) STP connector (4) Purge port Figure 4.1 Configuration of the STP Pump...
When you use gases including alkaline metals, but excluding Li, gases including Ga, Hg, In, or Sn, or HBr, contact BOC Edwards (See Section 1.1, "Usable Gases"). If the STP pump is used in an area with radiation, contact BOC ◇...
STP-H803/H1303 Series Instruction Manual 4.2.2 Installation Area Leave enough space for the following in addition to that for the STP pump: • Space for maintenance and inspection • Space for connecting cables The minimum bending radius of the STP connection cable is ◇...
Page 30
The bolt may not be able to be inserted from the lower side of the inlet port according to the shape of the inlet port flange. When the external appearance of the STP pump is not in the manual, contact BOC Edwards.
STP-H803/H1303 Series Instruction Manual How to Install the STP Pump Install the STP pump to the vacuum equipment as shown in Figure 4.2. Vacuum Equipment The STP pump can be installed through the damper (optional accessory). Inlet Port Inlet Port Flange Vacuum Valve Dry Pump, and other To Power Supply...
STP-H803/H1303 Series Instruction Manual 4.3.1 Cleaning the Seal Inspect the seals of inlet and outlet port flanges for dirt or oil spots before installing the STP pump to the vacuum equipment. Take the following measures for cleaning the seals: • Clean off with a pure gas.
STP-H803/H1303 Series Instruction Manual 4.3.2 STP Pump Installation Positions The STP pump can be installed vertically, horizontally, upside-down and slanted. Upside-down Horizontal Vacuum Equipment Slanted Vertical Figure 4.3 STP Pump Installation Positions When installing the STP pump in a horizontal or slanted position, it is recommended to install it so that the direction of the outlet port is on a vertical or horizontal plane in the direction of the gravity.
Use a material that has a tensile strength of 600N/mm or more. ◇ When securing the base, use stainless steel securing bolts with a tensile strength class is 70 or more. When using any securing method other than that specified in this ◇ manual, contact BOC Edwards.
STP-H803/H1303 Series Instruction Manual 1) When securing the inlet port with bolts 1) When securing the inlet port with bolts Refer to Table 4.2 for torque in pump abnormality and recommended securing bolts. Secure the inlet port flange with all of the boltholes of the size specified in the Inlet Port Flange Standard Reduced diameter shank bolts (R.D.S.B.) listed on Table 4.2 are more reinforced bolts over standard bolts by smoothing the portion to attach flange securing bolts to...
STP-H803/H1303 Series Instruction Manual Refer to the following Figure for the shape of R.D.S.B. Equipment Equipment Installation face Space by cantering r r Pump Flange Pump Flange VG/ICF Flange ISO_F Flange Bolt size Type of flange ISO_F flange 2.5 mm or more 5.9 mm or more ICF flange 1 mm or more...
STP-H803/H1303 Series Instruction Manual Vacuum Equipment Claw Clamps Secure the base Figure 4.7 Example of securing the STP pump Figure 4.7 Example of securing the STP pump (When securing the inlet port flange with claw clamps) 3) When installing the damper in the inlet port flange Refer to Table 4.2 for rotational torque.
STP-H803/H1303 Series Instruction Manual 4.3.4 Vacuum Piping DO NOT open the STP pump through the flange to atmospheric air ◇ while the STP pump is running. If atmospheric air flows into the STP pump, it may not function normally. Depending upon the type of the backing pump used, atmospheric air ◇...
Page 39
STP-H803/H1303 Series Instruction Manual 6) Depending upon the type of the backing pump used, oil vapor may 6) Depending upon the type of the backing pump used, oil vapor may contaminate the inside of the STP pump. Some oil viscosity could cause a contaminate the inside of the STP pump.
STP-H803/H1303 Series Instruction Manual 4.3.5 Connecting the Purge Port When pumping reactive or corrosive gases, introduce a dry N gas or other gas into the STP pump in order to protect the inside of the STP pump. As shown in Figure 4.9, introduce a dry N gas through the electromagnetic vent.
When you use gases including alkaline metals, but excluding Li, gases including Ga, Hg, In, or Sn, or HBr, contact BOC Edwards. Cool the STP pump to prevent the STP pump from overheating when ◇...
STP-H803/H1303 Series Instruction Manual Cooling the STP Pump When pumping gases, cool the STP pump. Also, when performing baking, always cool the STP pump. If the TMS unit is used, cool the STP pump with the method specified for the TMS unit.
STP-H803/H1303 Series Instruction Manual Baking the STP Pump To attain a less pressure in a shorter time and reduce the exhaust time, bake the vacuum equipment and STP pump. The surfaces of the STP pump and its peripheral equipment will ◇...
STP-H803/H1303 Series Instruction Manual Baking Heater Cooling Water Port PT(RC)1/4 Figure 5.1 Attaching Positions of the Cooling Unit and Baking Heater...
STP-H803/H1303 Series Instruction Manual Installation of the STP Control Unit Name and Function of Each Part 6.1.1 Front Panel (1) "START" Switch (flat panel switch, green) ・ STP pump rotating function (valid in MANUAL operation only). (2) "STOP" Switch (flat panel switch, dark gray) ・...
Page 46
STP-H803/H1303 Series Instruction Manual (9) LCD ・ The LCD displays a pump’s operation state, speed, or other messages. ・ The LCD displays an error message when an abnormality/error occurs in the STP pump. (10) "ACCEL." Lamp (green LED) ・ Lights during acceleration (ACCELERATION state). (11) "NORMAL"...
STP-H803/H1303 Series Instruction Manual (14) (15) STP CONTROL UNIT POWER FAILURE REMOTE (13) TEMP CTRL RESET SELECT NORMAL (16) ACCEL BRAKE DOWN ENTER START STOP (11) (10) (12) Figure 6.1 STP Control Unit Front Panel...
STP-H803/H1303 Series Instruction Manual 6.1.2 Rear Panel A hazardous live voltage may exist at connector that marked ◇ DO NOT touch the terminal. Doing so may result in electric shock. When operating connection/disconnection to connector, always power OFF the STP pump (Switch the breaker "OFF"). (17) AC POWER Connector (X2) A maximum voltage : Equal to the input voltage of this connector.
STP-H803/H1303 Series Instruction Manual Precautions Before Installation 6.2.1 Operating Environment Install the STP control unit in a place meeting the following requirements: Ambient Temperature 0 to 40 °C 30 to 95% (no dew condensing) Ambient Relative Humidity A place free of exposure to direct sunlight. •...
STP-H803/H1303 Series Instruction Manual 6.2.3 Installation Area Leave enough space for the following in addition to that for the STP control unit • Space for maintenance and inspection • Space for inlet and outlet of air for cooling - Top and side: 5 cm or more - Bottom: 1.8 cm or more (height of the rubber feet) •...
STP-H803/H1303 Series Instruction Manual Attaching the STP Control Unit to a Rack The dimensions of the STP control unit front panel conform to EIA standard. Therefore, this panel can be attached to any type of commercially-available racks. Attach the STP control unit to the rack according to the following steps: •...
STP-H803/H1303 Series Instruction Manual Cable Connection 6.4.1 Name and Dimensions of Each Cable Unit:mm STP Control Unit Side STP Pump Side Blue label(see NOTICE) 60 Pin ( Socket 60 Pin ( L-Type Connector Figure 6.5 External Dimensions of Each Cable There are two types of STP connection cables.
STP-H803/H1303 Series Instruction Manual Unit:mm STP Control Unit Side Power Side (Primary) Crimp-Type Terminal Lug (M4 ) 1.5mm ×3 cores 3 Pin (Socket) φ8.1 (Blown, Blue, Yellow/Green) or φ8.8 (Black1, Black2, Yellow/Green) N (L2) L (L1) PE( ) a) without power plug STP Control Unit Side Power Side (Primary) Plug type:NEMA6-15...
STP-H803/H1303 Series Instruction Manual 6.4.2 How to Connect the Cables Use the STP connection cable that BOC Edwards has specified. The ◇ use of other cables may result in product damage. When connecting/disconnecting cables, always power OFF the STP ◇...
STP-H803/H1303 Series Instruction Manual Table 6.1 Connecting the Power Cable Rear Panel CON1 Cable Color Remarks "AC POWER" Pin No. Outline of Connector Brown or L(L1) Black1 Single phase 200 to 240 V AC ±10% Both 50/60 Hz Blue or N(L2) Black2 ) Yellow/Green Ground...
STP-H803/H1303 Series Instruction Manual Connecting to Semiconductor Equipment The STP pump is a component system when installing to the semiconductor equipment. Consider the followings when designing the semiconductor equipment. 6.5.1 Connecting to Power The STP control unit receives its power from the semiconductor equipment electrical distribution system via a circuit breaker.
STP-H803/H1303 Series Instruction Manual How to Start/Stop the STP Pump NEVER connect or disconnect any cables while the power is ON. ◇ ◇ NEVER turn the primary power OFF (Switch the breaker "OFF") while the STP pump is in rotation. DO NOT release the inlet port flange or outlet port flange into the ◇...
STP-H803/H1303 Series Instruction Manual Start Procedures Start the backing pump before or simultaneously with start of the STP pump. Open the vacuum valve located at the outlet port flange side after starting the backing pump. DO NOT open the vacuum valve without operating the backing pump. ◇...
STP-H803/H1303 Series Instruction Manual There are two methods of operating the STP pump: MANUAL and REMOTE. Select one which fits your vacuum equipment. Manual Operation (See Table 7.1) Slide the "MANUAL/REMOTE" changeover switch on the STP control unit front panel to the position opposite to "ON." At this time, the "REMOTE" lamp is off. 7.4.1 Powering ON 1) Switch "ON"...
STP-H803/H1303 Series Instruction Manual 7.4.3 Stopping the STP Pump 1) Press the "STOP" switch on the front panel to stop the STP pump. The "NORMAL" lamp or "ACCEL." lamp goes out, and the "BRAKE" lamp lights (BRAKE state). 2) The "BRAKE" lamp goes out when the rotational speed goes below 500 rpm. 7.4.4 Starting the STP Pump after Stopping Press the "START"...
STP-H803/H1303 Series Instruction Manual Remote Operation (See Figures 7.1.) (Read through Section 8, "Remote Input/Output Signal Connector" before use.) Slide the "MANUAL/REMOTE" changeover switch on the STP control unit front panel to "ON." The "REMOTE" lamp lights. 7.5.1 Powering ON 1) Switch "ON"...
STP-H803/H1303 Series Instruction Manual 7.5.3 Starting the STP Pump after Stopping Perform the start operation to reaccelerate the STP pump. See Section 7.5.2, "Starting/Stopping the STP Pump." The STP pump can be reaccelerated even while it is stopping. 7.5.4 Powering OFF Switch "OFF"...
STP-H803/H1303 Series Instruction Manual Operating the TMS Unit (For use with the TMS unit only) 7.6.1 Starting/Stopping Power on the STP pump to automatically start the TMS unit. Power off the STP pump to automatically stop the TMS unit. While the TMS unit is in operation, the "TEMP CTRL" lamp of the STP control unit lights.
STP-H803/H1303 Series Instruction Manual <STP Pump State> <LCD Message> <Operation> Breaker ON Software Version Version v *.* 1997 ***** Autotest Self test in progress Execute Self Test (upper only) Self test completed Autotest complete (upper only) Levitation Upper:Levitation state Levitation State TMS TEMP : **℃...
STP-H803/H1303 Series Instruction Manual Remote Input/Output Signal Connector The remote input/output signal connector (REMOTE X7) is used for input/output remote signals. ∗ 1 This connector is of D-Sub type (37-pins, socket type). The screw for connector is M2.6 (19) (37) (20) REMOTE X7 Pin Arrangement Procure the connector for remote connection at your company.
STP-H803/H1303 Series Instruction Manual Table 8.1 REMOTE X7 Input Signal Pins Description Pins for inputting the START signal. The following two methods are available: 1) Short the circuits between (1)-(21). Then, short the circuits between (3)-(21) for 0.3 seconds or more. START IN However, when inputting this START signal simultaneously with switching "ON"...
STP-H803/H1303 Series Instruction Manual スタート1)の場合 In case of Start 1) スタート2)の場合 リモート入力信号 / Remote Input Signal In case of Start 2) START STOP リセット信号 RESET signal RESET 回転禁止信号 Rotation INHIBIT signal INHIBIT 未使用ピン Unused Pins Figure 8.1 REMOTE X7 Input Signal Pins 1 pin (0V) is insulated from the frame ground.
STP-H803/H1303 Series Instruction Manual Output Signal Pins Use output signal pins according to Table 8.2 and Figure 8.2. Remote output signals function during MANUAL and REMOTE operations. Four abbreviations are used in Table 8.2: N.O: Normal Open N.C: Normal Close COM: Common OUT: Output Pin Table 8.2 REMOTE X7 Output Signal Pins(1/2)
Page 69
STP-H803/H1303 Series Instruction Manual Table 8.2 REMOTE X7 Output Signal Pins(2/2) Description Pins for the TMS unit rated state signal output. AT TEMP. OUT (16) When the actual temperature of the TMS unit is within the setting temperature range ±10°C, the pins between (17)-(16) are closed and the pins between (35) (17)
STP-H803/H1303 Series Instruction Manual Normal Open : : Normal Close Remote Output Signals : Common R EM OT E 8 REMOTE selection state C R 2 signal output pins N .O O UT P OW ER 9 POWER ON state C...
STP-H803/H1303 Series Instruction Manual Table 8.3 shows the rated contacts for relays CR1 to 8 in Figure 8.2. Table 8.3 Rated Contacts for Relays CR1 to CR8 Resistance Load (COSφ=1) 125V AC 0.5A Rated Load 24V DC 1A Rated Current Maximum Contact Point Current Maximum...
STP-H803/H1303 Series Instruction Manual Adjustment Methods Tuning For the STP pump, tuning is required to align the position of levitation in the axial direction with the center of the rotor's movable range. Tuning can be performed simply by pressing the "RESET" switch on the front panel. Tuning is completed before delivery, so it is not required before use.
STP-H803/H1303 Series Instruction Manual Tuning can be performed only while the STP pump is in the ◇ LEVITATION state (It cannot be performed while the STP pump is in the ACCELERATION, NORMAL OPERATION, or BRAKE state). When performing the tuning during remote operation, use the ◇...
STP-H803/H1303 Series Instruction Manual <Contents> <Pump Operation> <LCD Display Example> Breaker "ON" Software Version v1.0 1997 ***** When the serial numbers Autotest do not match: Self Test Autotest Complete CAUTION Bad EEPROM content Levitation Levitation State Software Version Press the etc., a status is displayed.
STP-H803/H1303 Series Instruction Manual Confirmation Mode Confirmation Mode is used to check the status of the STP pump and STP control unit. The following items can be checked in Confirmation Mode: (1) Version information (2) Individual information (serial number, total hours of running, and number of starts) (3) Current settings (rated speed, TMS temperature setting, and acutual pump temperature)
STP-H803/H1303 Series Instruction Manual Parameter Set Mode Parameter Set Mode is used to set the different parameters of the STP pump and STP control unit. The following items can be set in Parameter Set Mode: (1) Remote mode (2) Rated speed (3) TMS function enable/disable (4) Inhibit function enable/disable (5) Emergency vent.
STP-H803/H1303 Series Instruction Manual ◇ Do not set the "RS232/485 Com", "RS232/485 Modbus", and "LONWORKS" settings of Remote Mode because they cannot be set in the current configuration. "UP" Switch :Change to the next value "DOWN" Switch :Change to the previous value "SELECT"...
STP-H803/H1303 Series Instruction Manual Remote Mode Detail The settings listed here are for "RS232/485 Com", "RS232/485 Settings Modbus", and "LONW ORKS" only. They are not displayed in "I/O Remote Mode." Baud Rate Baud Rate Baud Rate 9600 xxxx Press "SELECT" to go to next menu or "ENTER"...
STP-H803/H1303 Series Instruction Manual Safety Functions 10.1 Power Failure <Ⅰ. Operation at a Power Failure> When the power voltage drops below 80 V due to a power failure or other, the normal operation of the magnetic bearing is maintained using the regenerative energy of the rotating rotor (backup operation during a power failure).
STP-H803/H1303 Series Instruction Manual Establish a sequence so that the power can be supplied to the STP ◇ control unit immediately after a power recovery. Table 10.1 shows the states of LCD, LED lamps, and REMOTE output signals at a power failure. Also, Table 10.2 shows the operations of the STP pump after a power recovery.
STP-H803/H1303 Series Instruction Manual 10.2 Abnormal State of Magnetic Bearing When the magnetic bearing does not function normally due to a breakage of the STP connection cable, disconnection of connectors or any abnormality/error of the STP control circuit, the rotor falls on the touch down bearing and stops. The "FAILURE"...
STP-H803/H1303 Series Instruction Manual 10.6 Overheating Inside the STP Control Unit When the temperature inside the STP control unit overheats due to a failure in the air cooling fan, external heat source, and other, the WARNING message "CAUTION: CNT heat*" is displayed. In this case, the STP pump does not stop. If the STP control unit overheats further, the STP pump decelerates and stops.
Page 84
STP-H803/H1303 Series Instruction Manual 2) Disconnecting of the TMS Connection Cable and Connectors When an abnormality/error occurs in the TMS connection cable, the STP pump decelerates and stops. Output of the heater and the cooling water unit is OFF. The "FAILURE" lamp lights and the LCD displays error message "T.Cable Disconnect".
STP-H803/H1303 Series Instruction Manual Troubleshooting, Maintenance, and Inspection The STP pump is provided with safety functions for various abnormality/errors. When an abnormality/errors occurs in the pump or control unit, the safety function operates with the "FAILURE" lamp lit, and an alarm signal is output from the remote output pins.
STP-H803/H1303 Series Instruction Manual 11.2 When Error Message is Displayed on LCD 11.2.1 Error Messages When error is detected, "Error n˚ **/**" and the error message are displayed on the lower line of the LCD alternately. The first number of "**/**" is the order of detected errors that have occured.
STP-H803/H1303 Series Instruction Manual Table 11.1 Error Messages Referred LCD Messages Probable Causes Countermeasures Section CAUTION 1) Turning on the power after 1) Perform tuning. Bad PCB EEPROM "Tuning Error" is detected. 2) Abnormal data in STP 2) Perform tuning. control unit.
Page 91
STP-H803/H1303 Series Instruction Manual Referred LCD Messages Probable Causes Countermeasures Section DRIVER Overload The motor driver overload. (when the rated operation does not start approx. 30 min. after start operation is executed: ACCELERATION state continues for approx. 30 minutes: 1) High pressure at the inlet 1) Use the maximum 4.3.4 port...
Page 92
STP-H803/H1303 Series Instruction Manual Referred LCD Messages Probable Causes Countermeasures Section Error Message n°**/** If an error occurs Refer to the respective 11.2 (**/** shows the number of error message. errors occurred). First Damage Limit First damage phenomenon Overhaul is needed. Malfunction appeared on the touch down Contact Service office.
Page 93
STP-H803/H1303 Series Instruction Manual Referred LCD Messages Probable Causes Countermeasures Section P.Cable Disconnected Incorrect connection of the STP Connect the cable 6.4.2 connection cable. correctly. If the cable connector is not locked, lock it securely. Failure of the STP connection Contact Service office.
Page 94
STP-H803/H1303 Series Instruction Manual Referred LCD Messages Probable Causes Countermeasures Section TMS Higher Temp. TMS temperature control error (when the TMS unit detection temperature becomes +10°C or more): 1) Insufficient cooling of the 1) Connect the cooling Instruction TMS unit water Manual of electromagnetic...
STP-H803/H1303 Series Instruction Manual 11.3 Other Abnormalities 11.3.1 Abnormalities When Powering ON (When the "FAILURE" lamp lights and the LCD displays an error message, see Section 11.2, "When Error Message is Displayed on LCD.") Table 11.2 Troubleshooting after Powering ON Referred Symptom Probable cause...
STP-H803/H1303 Series Instruction Manual 11.3.3 Other Abnormalities (When the "FAILURE" lamp lights and the LCD displays an error message, see Section 11.2, "When Error Message is Displayed on LCD.") Table 11.4 Troubleshooting While the STP Pump Is Rotating Referred Symptom Probable causes Countermeasures Section...
STP-H803/H1303 Series Instruction Manual 11.4 "WARNING" Message Display Function The STP control unit is provided with a "WARNING" message to display when an overhaul is needed following a self test. When a "w" appears in the upper left hand corner of the LCD, press and hold the "ENTER"...
STP-H803/H1303 Series Instruction Manual 11.4.2 Second Damage Limit "Second Damage Limit" is displayed when continuing to operate the STP pump after "First Damage Limit" is displayed, and the accumulation of the damage point attains the second setting value by generating second "Disturbance" or "Mains Failure".
STP-H803/H1303 Series Instruction Manual 11.6 Maintenance and Inspection When performing maintenance and inspections of the STP pump, ◇ exhaust gases inside the STP pump thoroughly. Residual gases may cause an accident when removing the STP pump. Confirm the characteristics of gas to be used, referring to the Material Safety Data Sheet (MSDS) you obtain from the gas supplier.
STP-H803/H1303 Series Instruction Manual 11.6.1 Cleaning and Decontamination The method of cleaning and decontamination of the STP pump and STP control unit is shown below. Table 11.5 Cleaning and Decontamination Area Cleaning and Decontamination STP Pump Exterior Wipe with proper solvent (such as alcohol). Clean off with a pure gas when dust has accumulated in the connector.
Contact with atmospheric air may cause a reaction of the deposit. ◇ DO NOT open the STP pump to atmospheric air as much as possible. BOC Edwards supplies you the Temperature Management System ◇ (TMS) unit to keep the temperature high inside the STP pump and prevent the accumulation of deposit.
5 years (Only vacuum pumping) Other use (2 years) Dependent on application, contact BOC Edwards. The touch down bearing inside the STP pump will be worn out after a ◇ number of full speed touchdowns. When the STP control unit displays "w,"...
When returning the STP pump to Service office, be sure to pack it ◇ well to prevent external damage. If "Return Procedure" has not been satisfied, BOC Edwards will not be responsible for any troubles. i) Always contact Service office before returning the STP pump for repairs, overhaul, or other purposes.
STP-H803/H1303 Series Instruction Manual Storage and Disposal 12.1 Storage of the STP Pump When the STP pump is left unused over a long period (more than a few months), follow the precautions below: 1) Close the inlet port of the STP pump and vacuum it using a backing pump. 2) Introduce dry N gas or dry air from the outlet port or the purge port.
Confirm the characteristics of gas to be used, referring to the Material Safety Data Sheet (MSDS) you obtain from the gas supplier. BOC Edwards is not responsible for problems during or after ◇ disposal.
STP-H803/H1303 Series Instruction Manual Specifications 13.1 Specifications for the STP Pump Table 13.1 Specifications for the STP Pump Item H803 Series H1303 Series Inlet port flange VG150/ISO160 VG200/ISO200 Flange size Outlet port flange KF40 KF40 1300 Pumping speed 1000 >10 Compression ratio Ultimate pressure Pa (Torr)
STP-H803/H1303 Series Instruction Manual 13.2 Specifications for the STP Control Unit Table 13.2 Specifications for the STP Control Unit (1/2) Item Specifications Input voltage V AC 200 to 240±10% Without the TMS unit: Max. 850 Input power With the TMS unit: Max.
Page 111
STP-H803/H1303 Series Instruction Manual Table 13.2 Specifications for the STP Control Unit (2/2) Item Specifications Input/output terminals • AC POWER X2 ( 3 pins) • P.CONNECTOR X1 (60 pins) • TMS X5 (26 pins) • REMOTE X7 (37 pins) Input/output cables •...
Page 115
For more information, contact to the nearest Service Office. Manufacture: BOC Edwards Japan Limited 2-4-1, Shibakoen, Minato-ku, Tokyo 105-0011 JAPAN Telephone: Domestic 03-5470-6530 International +81-3-5470-6530 Facsimile: Domestic 03-5470-6521 International +81-3-5470-6521 Copyright 2003 BOC Edwards Japan Limited. All rights reserved.
Page 116
"STP-H803/H1303 Series Instruction Manual." Keep this Manual in a place where you can quickly access it at any time. BOC Edwards Japan Limited Dec. 2003 Copyright 2003 BOC Edwards Japan Limited. All rights reserved. Printed in Japan.
Page 117
STP-H803C1 Instruction Manual (excerpt) INTRODUCTION Thank you very much for purchasing BOC Edwards’ turbomolecular pump. The turbomolecular pump is designed to be installed in the vacuum equipment to exhaust gases from it. This manual covers all items necessary to ensure safe installation, operation and...
Page 118
STP-H803C1 Instruction Manual (excerpt) TABLE OF CONTENTS 1 Specifications...................1-1 1.1 Specifications for the STP Pump..................1-1 1.2 Specifications for the STP Control Unit ................1-2 TABLES Table 1.1 Specifications for the STP Pump ................1-1 Table 1.2 Specifications for the STP Control Unit ..............1-2...
Page 119
STP-H803C1 Instruction Manual (excerpt) Specifications Specifications for the STP Pump Table 1.1 Specifications for the STP Pump Item STP-H803C1 Inlet port flange VG150/ISO160 Flange size Outlet port flange KF40 Pumping speed >10 Compression ratio Ultimate pressure Pa (Torr) ) order (after baking) ∗...
Page 120
STP-H803C1 Instruction Manual (excerpt) Specifications for the STP Control Unit Table 1.2 Specifications for the STP Control Unit (1/2) Item Specifications Input voltage V AC 200 to 240±10% Without the TMS unit: Max. 850 Input power With the TMS unit: Max.
Page 121
STP-H803C1 Instruction Manual (excerpt) Table 1.2 Specifications for the STP Control Unit (2/2) Item Specifications Input/output terminals AC POWER X2 ( 3 pins) P.CONNECTOR X1 (60 pins) TMS X5 (26 pins) REMOTE X7 (37 pins) Input/output cables Power (primary) cable STP connection cable TMS connection cable (optional accessory) Safety functions...
Page 122
For more information, contact to the nearest Service Office. Manufacture: BOC Edwards Japan Limited 2-4-1, Shibakoen, Minato-ku, Tokyo 105-0011 JAPAN Telephone: Domestic 03-5470-6530 International +81-3-5470-6530 Facsimile: Domestic 03-5470-6521 International +81-3-5470-6521 Copyright 2003 BOC Edwards Japan Limited. All rights reserved.
Page 123
7. If the equipment is large, strap the equipment and its accessories to a wooden pallet. If the equipment is too small to be strapped to a pallet, pack it in a suitable strong box. 8. Fax or post a copy of the Declaration (HS2) to BOC Edwards. The Declaration must arrive before the equipment.
Page 124
Read the Return of BOC Edwards Equipment - Procedure (HS1) before you complete this Declaration • Contact BOC Edwards to obtain a Return Authorisation Number and to obtain advice if you have any questions • Send this form to BOC Edwards before you return your equipment...