1.15
Install a stem cap.
1.15.a
Obtain a stem cap from the cleaning kit. Place it on the stem on the
front of the cell and press firmly to slide it on as far as it will go.
Step 1.15.a
1.15.b
Gently pull on the stem cap to make sure that it is secure. If it moves
easily or comes off, discard the stem cap and obtain another one from
the kit.
CAUTION
1.16
Verify that the cell is ready for cleaning:
•
Pills removed
Pleated breather removed from the top of the cell
•
Foam breather removed from the nozzle cover
•
•
Nozzle cover replaced
•
Cell door closed
Baffle covers removed (one from each side of cell)
•
Drug label removed (if applicable)
•
•
Splash guard removed
Adhesive residue removed from all areas on the cell
•
Cap placed on the stem
•
©2015 Parata Systems, LLC
120-0465 Rev 03
The cap must fit snugly to prevent liquid from getting into the
cell.
page 14 of 40
Customer Guide
HPAC Cell Cleaning
Parata Technical Assistance Center
1-888-PARATA1 (727-2821)