Width adjustment of footrests
If the footrest width needs to be adjusted, please undo the screw
(1), set the desired width, by positioning 1, 2 or 3 spacers (2) from
outer to the inner side and then refit the screws, (Fig.6.3.5).
Lightweight footrest
The footboard angle can be adjusted by loosening the screws (2).
Side protection panels can be fitted to the footboard using the
mounting holes (3) These prevent the feet slipping off accidentally,
(Fig.6.3.6). Make sure that after any adjustment work, all screws
are tightened correctly (see the page on torque).
Width adjustment
To adjust the width of the footboard open the screw (2). Then you
can slide in and out the tubes in the holder to adjust the width of the
footboard (Fig.6.3.6).
Please make sure that the screws are tightened with the correct
torque .
Automatic folding footrest
The footrest angle can be adjusted by loosening the screws (1) and
turning the footrest. When tightening, please note the torque
measurements, (Fig.6.3.7).
High-mount footrest
The high-mount footrest is fitted on the inner part of the frame and
permits a higher position than a standard footrest (Fig.6.3.8)